2.5 Electrical Interconnection
The Series 10D1477 CK-MAG magnetic
f l o w m e t e r i s f u rn i sh e d w i t h a re m o t e
mounted signal converter. Interconnection
details are provided in the instruction man-
ual supplied with the signal converter.
Interconnection wiring is done in the Cus-
tomer Connection box using the terminal
blocks mounted on the circuit board. This is
shown in Figure 2-13 below.
WARNING
Equipment powered by ac line voltage
constitutes a potential electric shock
hazard to the user. Make certain that
the system power input leads are
disconnected from the operating
branch circuit before attempting
electrical interconnections.
Regardless of which system is used, the
grounding procedures given in Section 2.4,
must be followed.
*
For applications other than
ANSI CLASS 150, contact the manufacturer .
FIGURE 2-13. CUSTOMER CONNECTION BOX FOR
INTERCONNECTION WIRING
FIGURE 2-12. OUTLINE DIMENSIONS,
GROUNDING RINGS*
10D1477 MAGNETIC FLOWMETER INSTRUCTION MANUAL