4.4 Jumper Selections
4.4.1 Model 10D1475S
For the integral-mounted converters, zero return and functions associated with the 50XM1000 signal
converter are established by the movable jumpers at terminals A2 and B1 located on the lower left
corner of the
686B623U01 Primary Board
assembly. The following table and diagram correlates the
zero return and functions with the jumper connections:
Jumpers
Functions
A2
B1
Zero return
ZR
G3
10 kHz
8D
G3
Pulse reverse
1R
9R
4.4.2 Model 10D1475J
4.4.2.1 Integral
For the integral version of Model 10D1475J Converter, zero return and functions associated with the
50XM1000 signal converter are established by the movable jumpers on terminal-strip J1 located in
the lower left corner of the
686B762U02 Primary Board
assembly. The following table and diagram
correlates the zero return and functions with the jumper connections:
Jumpers
Functions
Jumper
Position
Zero return
1-2
10 kHz
3-4
Pulse reverse
5-6
The full view of the 686B762U02 Primary Board Assembly is shown in Figure 4-2 along with a table
of selectable functions and their jumper settings.
10D1475 MINI-MAG MAGNETIC FLOWMETER INSTRUCTION MANUAL
4-6