Troubleshooting
(Continued)
Abaxis Technical Support can be reached as the system administrator.
Quality Check Failure Resolutions
Resolutions (Continued)
with a freshly filled cartridge. Carefully observe the help provided
throughout the testing pathway. If the same quality check failure displays,
contact the system administrator for further instruction.
L
A cartridge was detected when an Electronic Simulator was expected.
Repeat testing ensuring to insert an Electronic Simulator. Make sure that
the cover retaining ring does not interfere with Simulator insertion. Ensure
the Simulator is fully inserted. The Simulator is fully inserted when the click
is heard. If the same quality check failure displays, contact the system
administrator for further instruction.
M
Reserved for future use.
N
The cartridge was not fully inserted. Repeat testing with a freshly filled
cartridge. Ensure the cartridge is fully inserted. The cartridge is fully
inserted when the click is heard. If the same quality check failure displays,
contact the system administrator for further instruction.
P
The analyzer did not reset correctly. Repeat electronic simulator testing.
If the test results in a PASS, the analyzer is ready for use otherwise
contact the system administrator for further instruction.
R
Test has been successfully cancelled.
S
OSi software installation required. Contact the system administrator for
further instructions
T
Install the most recent OSi software. Contact the system administrator for
further instructions.