Disinfecting
Maintenance
(Continued)
i-STAT
Alinity v
— User Manual
650-7100 Rev. A
41
i-STAT Alinity v Analyzer
The disinfection process must begin
IMMEDIATELY
after the cleaning procedure is
complete. Standard precautions should be taken whenever working with blood or
blood products.
Remove a wipe from the container
and squeeze out any excess fluid
before you begin to wipe the
surface.
Wipe all surfaces three times.
This includes front, back, sides,
top and bottom.
WIPE
WIPE
3X
3X
WIPE
WIPE
3X
3X
W
I
P
E
3
X
3
X
Wipe with gauze
until dry.
Allow all surfaces
to remain wet
for
3 minutes
.
Sensitive Areas
Avoid forcing liquid into these areas:
i-STAT Alinity v Analyzer
Cartridge Port
10-Pin Connector under the camera
Gold contacts (2) on the outside of the battery
B
A
C
2
1
3
4
A
B
C