35
8.5 Technical safety test
WARNING - Technical safety tests may only be performed by the manufacturer, the AAT field
service or an authorised specialist dealer!
WARNING – It is strongly recommended to perform a technical safety test every 2 years! The
test badge shows when the test is to be carried out at the latest! Tip: After a change in operator/
reuse!
8.6 Reuse (change in operator)
The term reuse is used when the AAT product of patient X is to be used for another patient Y.
Please contact the manufacturer, the AAT field staff or the authorised specialist shop.
The CR23 is suitable for reuse. A technical safety test and cleaning/disinfection work should be per-
formed prior to each reuse. (Chapters 8.3 and 8.5)
i
NOTE - You may not mend or bypass a damaged fuse!
Only new fuses with
40 A
may be replaced!
i
NOTE - If the fuse blows frequently, contact the manufacturer or an authorised specialist dealer!
i
NOTE - Only fuses purchased from the specialist dealer or AAT may be used as replacement
fuses!
Remove the fuse cover on
the battery pack. To do this,
loosen the screw.
Remove covers and old fuses
and insert new fuses.
» Replacement fuses are
available from the manu-
facturer!
Position fuse cover and fix it
with the screw.
8.4 Replacing fuses
To ensure flawless device function, we strongly recommend carrying out a tech
-
nical safety test every 2 years. We also recommend the technical safety test
prior to each reuse.
Training is required in order to be able to perform the technical safety test. For
more information, please visit: www.aat-online.de