Technology
c-max U2 C1612
Dimensions
Refer to chapter "Dimensional drawing"
Own weight
Stair climber c-max U2 incl. both armrests = 30.9 kg
Battery pack = 4.0 kg, Total weight without battery pack= 30.9 kg
Total permissible weight
194 kg
Maximum person weight (max. load)
160 kg
Climbing speed
1)
9 - 23 steps / minute (continuously adjustable)
Range per battery charge for an average
person weight of 75 kg
1)
Approx. 25
floors (accounts for
approx. 375 steps)
Range per battery charge for maximum per-
son weight of 160 kg
1)
Approx. 9 floors (accounts for approx. 135 steps)
Max. height of step
210/ 225 mm (with climbing height extension)
Min. depth of step
2)
140 mm
Required stair width (straight stairs)
Width of c-max + 100 mm
Min. platform size (U-shaped stairs)
1350 x 1350 mm
Min. radius of spiral stairs
1350 mm
Min. depth of step (spiral stairs)
2)
140 mm
Max. slope of the stairs
60.3°
Tilt stability
In direction of travel 3.7° / against direction of travel 2.7°
Batteries
2 x 12 V / 5 Ah (gas-tight)
Direct current motor
24 V / 275 W
Switching force UP/DOWN switch
3-8 N
Direct operating forces
Max. 350 N when climbing stairs - downwards (chapter 8.5)
Sound pressure level
< 67 dB
Storing/operating temperature: Device
-40 °C to +65 °C / -25 °C to +50 °C
Storing/operating temperature: Battery
-20 °C to +65 °C
(3)
/ -25 °C to +50 °C
(3)
Battery charging temperature
0 °C to +40 °C
Altitude application range (sea level)
< 3000 m above sea level
Vibration information
The weighted RMS value of acceleration, to which the upper body is sub-
jected, is typically below 2.5 m/s
2
.
Adhesive label with serial number
On the left of the housing and rear side of the user manual
IP protection class
IPX4 degree of protection: protection against splashing water from all direc-
tions
Classification according to ISO 7176-28:2012
Type C (according to Annex A, Table 1)
1) With activated ComfortStep, the climbing speed and the range are reduced. The range also depends on the weight of the occupant.
2) Any steps of a smaller depth may not be travelled as in this case it is no longer possible for the wheel to be set down safely.
3) At temperatures below 0 °C the performance of the battery pack will decrease; therefore we recommend a storage temperature of at least 0 °C.