33
Interval measurement
1. Set the cycle to a value other than "OFF".
2. "CYCLE:
∗∗∗
" appears at the bottom of the home display and starts the interval measurement.
3. Blood pressure starts to be measured at the exact-time synchronized with the built-in clock.
4. To end the interval measurement, set the cycle to "OFF".
Note
•
During interval measurement, pressurization is performed by "AUTO".
•
While in the interval measurement mode, blood pressure can be measured at any time by pressing the
NIBP START/STOP key.
•
When the notice buzzer is set to "ON", the buzzer sounds 15 seconds before the interval measurement
starts.
Continuous measurement
1. Set the cycle to "CON".
2. "CYCLE: CON" appears at the bottom of the home display and starts the continuous measurement
after 10 seconds.
3. After 5 minutes, the continuous measurement switches to the 5-minute interval measurement, and
the display switches from "CYCLE: CON" to "CYCLE: C-5".
4. To perform the continuous measurement again, go to the setting mode to set the cycle to "CON".
Quick systolic display
During interval measurement, the predicted systolic blood pressure flashes. It is replaced with the
measured value after measurement
6-6 Measuring the ECG
1. Turn on the POWER switch to start the ECG measurement.
2. Set the conditions appropriate for measurement.
Note
•
The instrument performs real-time detection of the QRS wave, measures the RR interval, and converts
it into the heart rate.
•
It may take about 10 seconds to display the heart rate after attaching the ECG electrodes.
•
Some patients have a weak R-wave pulse. In that case, change the induction method.