27
Applying the SpO
2
sensor
Use the sensor appropriate for the patient.
Refer to the instruction manual for each SpO
2
sensor for how to apply the sensor.
Warning
•
Read the SpO
2
sensor instruction manual for proper use.
•
Applying the sensor on a body part other than specified or using it longer than
specified may cause the patient to suffer physical injury.
Caution
Do not apply the SpO
2
sensor too tight.
Note
•
Do not apply the sensor on the same arm with an arterial catheter inserted or the cuff attached. Using
the same arm will lower the reliability of the measurement data.
Attaching the temperature probe
The temperature probe and the temperature receiver unit are optional. The temperature receiver unit is to
be connected to the OPTION connector.
For how to use them, refer to the relevant instruction manuals.
Caution
The temperature probe that can be used with the instrument is of the wireless type.
The signal reception condition depends on the environment. Keep it away from
electrical noise sources.