17
Name
Displayed
item
Description
Systolic blood pressure
•
Displays the value after measurement.
•
Retained until the next measurement.
Systolic blood pressure
(When an alarm occurs)
•
Flashes in red when the limit values are exceeded.
•
Retained until the next measurement.
Blood pressure error
•
Flashes the error code when an error occurs.
•
Retained until the next measurement.
Quick systolic
•
Flashes the predicted value during interval
measurement.
•
The value is not monitored by the limit function.
•
Replaced with the final value after measurement.
"SYS"
•
Displayed at the upper left side to indicate that the
systolic blood pressure value is displayed.
1
Systolic blood
pressure
display
" " (Red circle)
•
Displays a red circle near "SYS" when the limit
values are set.
Diastolic blood pressure
•
Displays the value after measurement.
•
Retained until the next measurement.
Diastolic blood pressure
(When an alarm occurs)
•
Flashes in red when the limit values are exceeded.
•
Retained until the next measurement.
Cuff pressure
•
Displays the cuff pressure value during
measurement.
"DIA"
•
Displayed at the upper left side to indicate that the
diastolic blood pressure value is displayed.
2
Diastolic blood
pressure
display
" " (Red circle)
•
Displays a red circle near "DIA" when the limit
values are set.
Mean blood pressure
•
Displays the value after measurement.
•
Retained until the next measurement.
3
Mean blood
pressure
display
"MAP"
•
Displayed near the value to indicate that the mean
blood pressure value is displayed.
Heart rate/Pulse rate
•
Displays as specified in the function setting. (Refer
to "6-14 Function Setting".)
When "AUTO" is selected, displays the value
according to the order of precedence of ECG, SpO
2
and blood pressure.
•
When no values are detected, displays the ECG
status.
Displays " - - - " during pulse detection, and "LF"
when the electrode is off the proper site.
"HR" / "PR"
•
Displays "HR" to indicate that heart rate is
displayed in ECG measurement.
•
Displays "PR" to indicate that pulse rate is
displayed in SpO
2
or blood pressure measurement.
Heart mark
•
Flashes in synchronization with the heart rate/pulse
rate.
Heart rate/Pulse rate
(When an alarm occurs)
•
Flashes in red when the limit values are exceeded.
•
Retained until the value is within the limit values.
"bpm"
•
Displayed as the unit of heart rate/pulse rate.
" " (Red circle)
•
Displays a red circle near "HR" or "PR" when the
limit values are set.
4
Heart rate
Pulse rate
display
Displaying color
•
Displays the heart rate by the ECG measurement in
green.
•
Displays the pulse rate by the SpO
2
measurement
in blue.
•
Displays the pulse rate by the blood pressure
measurement in yellow.