25
6. Operation
6-1 Turning on the POWER switch
Turn the POWER switch on. After several seconds of self-check, the home display appears.
Note
•
When the self-check detects any error, the error code is displayed. The error code display of some
errors is frozen. Refer to "7-6 Error Codes" for how to solve the error.
•
The self-check checks the data in memory. The duration of the self-check depends on the number of
memory data.
6-2 Attaching the Cuff/ECG Electrodes/SpO
2
Sensor
Consider "Safety Precautions" appearing in an earlier part of the instruction manual, when attaching the
Cuff/ECG electrodes/SpO
2
sensor.
Attaching the cuff
1. Hold the cuff with the hose pointing
toward the lower part of the arm.
2. Wrap the cuff around the upper arm,
about 3 5 cm above the elbow, with the
ARTERY mark on the artery, as shown in
the illustration.
3. Allow some slack between the cuff and
arm so that a finger or two can be
inserted.
4. When the cuff inner end is not within the
area labeled RANGE, the cuff size is not
appropriate. Replace the cuff.
Note
•
Constriction of the upper arm, caused by rolling up a shirt sleeve, may prevent accurate readings.
•
When the size of the supplied cuff does not fit the patient, refer to "8-1 Accessories/Options" to find the
proper sized cuff.