47
7-4-4 RsVisco Format
Used with the graphing program RsVisco. Measurement data are output in comma separated format.
When a comma is selected as the decimal point by "
pnt 1
", a semicolon ";" is used as a data separator.
When a measurement is started using RsVisco, the viscometer automatically selects this format.
Measurement data are output in the order of viscosity, viscosity unit, temperature and temperature
unit.
The output data are 25 characters long excluding the terminator
With RsVisco format selected, the viscosity value and the temperature are output using the internal resolution.
The relation between the measuring unit and the internal resolution is as follows:
Viscosity Temperature
Model
mPa
・
s Pa
・
s cP P
°
C
°
F
SV-10 0.01 0.0001 0.01 0.0001
Internal
resolution SV-100
- 0.01 - 0.1
0.01 0.01
SV-10 output format example
Viscosity
/
Temper-
ature
Display
Output format example
Remarks
L
LLLL
mPa
・
s
+00000.00,mPa s,+025.67,C
Zeroes are output for below measuring
range error.
0.30 mPa
・
s
+00000.30,mPa s,+025.67,C
10.0 mPa
・
s
+00010.00,mPa s,+025.67,C
100 mPa
・
s
+00100.00,mPa s,+025.67,C
1.00 Pa
・
s
+01000.00,mPa s,+025.67,C
For 1000 mPa
⋅
s or greater, the displayed unit
is Pa
⋅
s, but the output unit remains mPa
⋅
s.
mPa
・
s
/
°C
H Pa
・
s
+12000.00,mPa s,+025.67,C
12000 is output for above measuring
range error.
L Pa
・
s
+000.0000, Pa s,+051.23,F
Zeroes are output for below measuring
range error.
0.0003 Pa
・
s
+000.0003, Pa s,+051.23,F
0.0100 Pa
・
s
+000.0100, Pa s,+051.23,F
0.100 Pa
・
s
+000.1000, Pa s,+051.23,F
1.00 Pa
・
s
+001.0000, Pa s,+051.23,F
Pa
・
s
/
°F
H Pa
・
s
+012.0000, Pa
,+051.23,F
12 is output for above measuring range error.
L cP
+000.0000, cP
,+025.67,C
Zeroes are output for below measuring
range error.
0.30 cP
+00000.30, cP
,+025.67,C
10.0 cP
+00010.00, cP
,+025.67,C
100
cP
+00100.00, cP
,+025.67,C
10.0 P
+01000.00, cP
,+025.67,C
For 1000 cP or greater, the displayed
unit is P, but the output unit remains cP.
cP
/
°C
H P
+12000.00, cP
,+025.67,C
12000 is output for above measuring
range error.
L
P
+000.0000,
P
,+051.23,F
Zeroes are output for below measuring
range error.
0.0030 P
+000.0030,
P
,+051.23,F
0.100 P
+000.1000,
P
,+051.23,F
1.00 P
+001.0000,
P
,+051.23,F
10.0 P
+010.0000,
P
,+051.23,F
P
/
°F
H P
+120.0000,
P
,+051.23,F
120 is output for above measuring range error.
: Space (ASC 20h)