The stability of the viscosity measurement results can be adjusted, taking ambient conditions such
as vibration into consideration.
Parameter
Settings Description
0
Follows the viscosity changes
quickly.
(Prone to vibration)
1
•
2
Follows the viscosity changes
slowly.
(Stable values)
When the viscosity value is unstable due to external
vibration, set a greater parameter.
To measure while following the rapid changes in
viscosity, set a smaller parameter.
With a smaller setting, the measurement is prone to
external vibration. Consider the ambient conditions
of the installation site.
Unit Upon Power-on (
Unit
)
The units of viscosity and temperature displayed when the power is turned on are specified.
SV-10
Parameter
Settings Description
0
•
mPa
⋅
s
(Millipascal second)
1
Pa
⋅
s
(Pascal second)
2
cP
(Centipoise)
3
P
(Poise)
°C
(Celsius)
4
mPa
⋅
s
(Millipascal second)
5
Pa
⋅
s
(Pascal second)
6
cP
(Centipoise)
7
Viscos-
ity
P
(Poise)
Temper-
ature
°F
(Fahrenheit)
In the standby mode, pressing
the MODE key switches the
viscosity unit.
mPa
⋅
s
⇔
Pa
⋅
s, cP
⇔
P
With "
fnc 0
" selected, units can
be switched even during
measurement. *1
SV-100
Parameter
Settings Description
1
•
Pa
⋅
s
(Pascal second)
3
P
(Poise)
°C
(Celsius)
5
Pa
⋅
s
(Pascal second)
7
Viscos-
ity
P
(Poise)
Temper-
ature
°F
(Fahrenheit)
In the standby mode, pressing
the MODE key switches the
viscosity unit.
Pa
⋅
s
⇔
P
With "
fnc 0
" selected, units can
be switched even during
measurement. *1
*1 While the measurement is being performed using the graphing program RsVisco, unit
changes using the MODE key is not available.
With the SV-10, for a viscosity over 1000 mPa
⋅
s, the unit is fixed to Pa
⋅
s, and for a
viscosity over 1000 cP, the unit is fixed to P.