7. FUNCTION SETTING
The viscometer, by selecting functions to be used in the function setting, can specify the
performance appropriate to the usage.
Each function is assigned parameters. The performance of a function is specified by changing the
parameter.
The parameters saved, even if the power is turned off, are maintained in non-volatile memory.
7-1 Operation
The operational procedure of the function setting is as follows:
1 In the standby mode, press and hold the MODE key to enter the function setting mode.
2 Press the MODE key to select a function item.
3 Press the PRINT key to confirm the function item. The changeable digit blinks.
4 Press the START key or HOLD key to change the blinking digit.
START key
Increases the value of the blinking digit. When the value reaches the upper
limit of the setting range, the minimum value appears again.
HOLD key
Decreases the value of the blinking digit. When the value reaches the lower
limit of the setting range, the maximum value appears again.
5 To save the new setting, press the PRINT key. After "
end
", the next item is displayed.
To cancel the new setting, press the STOP key. The next item is displayed.
6 To change other settings, repeat the procedure starting at step 2.
7 To exit the function setting mode, press the STOP key. The viscometer returns to the
standby mode.
Note
The operational procedures for setting the date and
time ("
Cladj
"), ID number ("
id
") and initialization
("
Clr
") are not the same as the procedure described
above. Refer to "7-3 Description of Items".
"Date/Time" pages
31-33
"Device ID Number"
pages 37-38