AD-4407A Weighing Indicator
Page 55
13.4.
UFC Command
UFC (Universal Flexi Coms) function enables editing the serial data output
format freely using the serial interface command.
For customizing the printout of the printer or efficient data collection.
Output data is not only the indicator’s data/status but also characters at will.
It can output the control code* of the printer. (* depends on the individual printer)
There are 2 set of memories for storing the parameters.
UFC Command Parameter
UFC commands such as SF1 have many parameters.
One command line can have a multiple number of parameters. The parameters
are stored in memory in order.
Multiple UFC commands are possible. The parameters are stored next to the
last parameter stored by the last UFC command.
Clear all of the data first, if storing a new set of parameters. The parameters in
the stored data can not be changed partially.
The various types of parameters and their descriptions are shown below.
data
Weight, result of comparison etc.
$CL
Clear previous settings. UFC command parameters can not be
changed partially.
$WT
Displayed data
$GR
GRoss data
$NT
NeT data
$TR
TaRe data
$HD
HeaDer of Gross/Net/Tare or Preset tare. Refer to CF06.
$UT
Weight UniT
$ST
STable/Unstable
$CP
Result of ComParison
$ID
ID number specified at F06
$DN
Data Number increments with each output automatically
$CD
CoDe memory number
$AN
Accumulation count
$TL
TotaL weight
$CM
CoMma
$CR
CR code (0Dh)
$LF
LF code (0Ah)
$DE
DElete the last parameter
$DL
Inserting DeLay time (0.1 second step)
Example: DL10: 1.0 second delay
13.4. UFC Command Parameter
Summary of Contents for OP-03
Page 65: ...AD 4407A Weighing Indicator Page 65 MEMO ...
Page 66: ...Page 66 AD 4407A Weighing Indicator MEMO ...
Page 67: ......