HW-G Instruction Manual Page 48
HW-G OZIM v1a
16 RS-232C Serial Data Interface
RS-232C interface has the following two modes.
Stream mode
Outputs data countinuously and can be used for printing data.
Command mode Controls the scale using commands from a computer.
It is necessary to set the parameters of the “Baud rate (
f4
)”,
”Output mode (
f5
)”, “ Format (
f15
)” in the function table in advance.
There are option cables as follows:
AX-KO557A RS-232C cable, D-sub 25 pin, 2m
AX-KO1786-200 RS-232C cable, D-sub 9 pin, 2m
Transmission system
EIA RS-232C
Transmission
Asynchronous, bi-directional, half-duplex
Data format
Baud rate: 2400, 4800, 9600 bps
Data:
7 bits
Parity:
1 bit, Even
Start bit
1 bit
Stop bit
1 bit
Code
ASCII
Terminator
CR LF (CR: 0D[H], LF: 0A[H])
Pin Connections
Pin No. Signal name
Direction Description
1
-
2
RXD
Input
Receive data
3
TXD
Output
Transmit data
4
-
5
SG
-
Signal ground
6
-
7
DSR
Output
Data set ready
Circuit
Summary of Contents for HW-G Series
Page 1: ......
Page 2: ......
Page 68: ...HW G Instruction Manual Page 66 HW G OZIM v1a Dimensions...
Page 69: ...HW G Instruction Manual Page 67 HW G OZIM v1a...
Page 71: ......