HW-G Instruction Manual Page 37
HW-G OZIM v1a
Caution
The parameters of the upper limit value (HI) and a final value (HI) use the same
memory. The parameters of the lower limit value (LO) and the zero band (LO) use
the same memory.
The upper/lower comparator function, the simple batch function and the full/dribble
batch function cannot be used at the same time because these parameters use
common memory.
13.1 Preparation (Setting Parameters)
Step 1 Turn off the display.
Press the
ON/OFF
switch while the
ZERO
switch
is pressed and held. The function table is displayed.
Step 2 Press the
ENTER
switch to display an item of the
comparator function (
f6
).
Step 3 Select a parameter of the simple batch function
(
f6 8
) with the
^
switch.
Step 4 Press the
ENTER
switch to store the new
parameter.
Step 5
Press the
F
switch and the
ENTER
switch to exit
from the function table. Then the scale displays the
weighing mode.