Page 60
HV/HW-C/CP Series
16.2. HVW-02CB, USB Interface
The HVW
-
02CB enables duplex communication using a USB interface.
The connector is type B.
To connect the HVW
-
02CB to a computer, a commercially "type A male
-
type B female"
cable can be used.
To communicate between the scale and a computer using the USB interface, it is necessary to
install the specified driver software to a computer. Download the driver software from the A&D
website.
The HVW
-
02CB can only be connected with a computer.
Do not use a USB terminal as power supply terminal. Do not connect a device other than a
computer to the USB terminal. Dosing so may cause failure and malfunctions.
16.2.1. Procedure for Using the USB Interface
Prepare the computer as shown below.
Step 1 Download the USB driver software from the A&D website.
Step 2 Install the USB driver software to the computer that is used as a COM port.
Step 3 Read the COM port number on the computer.
Step 4 Input the number to the data communication software.
Step 5 Connect the scale to the computer using the USB cable.
Step 6 When the PRINT key is pressed, a weighing value can be received by the data
communication software.
Example of data communication software
:
RsCom
of A&D
WinCT
.
16.2.2. Example of Using the USB Interface
Reading COM port number
( Computer is used as a COM port.)
Step 1 Download the USB driver software from the A&D website.
Open the A&D website at the URL
http://www.aandd.jp/
and
open the "
Support
" and "
Software
" pages in that order.
Download the driver software from "
download
" of "
Driver software for HVW-02CB
(USB
interface option for the HV/W
-
C & HV/W
-
CP series)".
Note
:
The location and version of the software may be changed without notice.
Step 2 Expand the file "
hvw-02cbja_driver.zip
" by double
-
clicking.
The folder "
hvw-02cbja_driver
" is created.
Step 3 Install the file "
CDM21224_setup.exe
" in the folder by double
-
clicking.
Step 4 Finish the installation by following the displayed instruction.