44
Interval output mode
The weight data is periodically output.
When the PRINT key is pressed, the balance starts to output the weight data at a preset interval
time. When the PRINT key is pressed again, the balance stops outputting the weight data.
Example
“For outputting the weight data periodically."
Required setting
dout prt
6
Interval output mode
dout int
Interval
time
Caution
The balance may not transmit the data completely at the specified interval times and baud
rate. Set the baud rate higher.
10-6 Description of the Item “Data Format”
A&D standard format
5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B
printer is used, set the printer to MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.
The unit, consisting of three characters, follows the data.
Output example:
Print example:
DP (Dump print) format
5if type 1
This format is suitable for the peripheral equipment that prints the received data as is.
When an AD-8121B printer is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
The unit, consisting of three characters, follows the data.
Unit
Data
Header
S T , + 0 1 2 .
7 8
g
CR LF
Terminator
0 0
Terminator
W T
+ 1 2 .
7 8 g
CR LF
Header
Data
Unit
0 0
Terminator
W T
+ 1 2 .
7 8 g
CR LF
Header
Data
Unit
0 0
Summary of Contents for HR-250AZ