76
21. SPECIFICATIONS
HR-250AZ HR-251AZ HR-150AZ HR-100AZ
Weighing capacity
252 g
252 g
62 g
152 g
102 g
Maximum display
252.0084 g
252.008 g
62.0009 g
*1
152.0084 g
102.0084 g
Minimum weighing value
(1 digit)
0.1 mg
1 mg
0.1 mg
0.1 mg
Repeatability
(Standard deviation
0.2 mg/200-250 g
0.1 mg/0-200 g
0.5 mg
0.1 mg
0.1 mg
Linearity
±0.3 mg
±1 mg
±0.3 mg
±0.2 mg
Stabilization time
(typical at FAST)
Approx. 2 second
*2
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F )
±2 ppm/°C
Internal mass
Yes
Clock and calendar function
Yes
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate
5 times/second
*3
, 10 times/second
Minimum unit mass
0.1 mg
1 mg
0.1 mg
Counting
mode
Number of samples
10, 25, 50 or 100 pieces
Minimum 100%
reference m-ass
10.0 mg
100 mg
10.0 mg
Percent
mode
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
Interface
RS-232C
External calibration weight
250 g
200 g
100 g
50 g
250 g
200 g
100 g
50 g
20 g
150 g
100 g
50 g
100 g
50 g
Weighing pan diameter
90 mm
External dimensions
198(W) x 294(D) x 315(H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Net weight
Approx. 3.9 kg
*1
: The balance allows weighing using a precision range, even with a heavy tare placed on the pan.
(Smart range function)
*2
: With MID. , factory setting of the rate of the response adjustment, the stabilization time is approx. 3 seconds.
*3
: Factory setting