48
10-9 Clock and Calendar Function (Only for the HR-AZ series)
The HR-AZ series balance is equipped with a clock and calendar function. When the “GLP output
(
info
)” parameter is set to “
1
” or “
2
” and the “Time/Date output (
5-td
)” parameter is set to “
1
”, “
2
” or
“
3
”, the time and date are added to the output data. Set or confirm the time and date as follows:
Operation
1 Press and hold the SAMPLE key until
ba5fnc
of
the function table is displayed.
2 Press the SAMPLE key several times to display
Cl adj
.
3 Press the PRINT key. The balance enters the mode to
confirm or set the time and date.
Confirming the date
4 The current date is displayed with all the digits
blinking.
When the date is correct and the operation is to be
finished, press the CAL key and go to step 8.
When the date is correct and the time is to be
confirmed, press the SAMPLE key and go to step 6.
When the date is not correct and is to be changed,
press the RE-ZERO key and go to step 5.
Note
The year is expressed using a two-digit format. For
example, the year 2010 is expressed as “10”.
Setting the date (with part of the digits blinking)
5 Set the date using the following keys.
SAMPLE key
To select the digits to change
the value. The selected digits
blink.
RE-ZERO key
To increase the value by one.
MODE key
To decrease the value by one.
key
To store the new setting,
display
end
and go to step
6.
CAL key
To cancel the new setting and
go to step 6.
Confirm
To the time
setting mode
To the time
setting mode