background image

 

48

Confirming the time 

6 The current time is displayed with all the digits 

blinking.  

  

When the time is correct and the date does not 

need to be confirmed, press the CAL key and go to 
step 8. 

  

When the time is correct and the date is to be 

confirmed, press the SAMPLE key and go back to 
step 4. 

  

When the time is not correct and is to be changed, 

press the RE-ZERO key and go to step 7.   

Setting the time (with part of the digits blinking) 

7  Set the time in a 24-hour format using the following 

keys.  

SAMPLE key 

To select the digits to change 

the value. The selected digits 
blink.

 

RE-ZERO key 

To increase the value by one.   

MODE key 

To decrease the value by one. 

PRINT 

key 

To store the new setting, 

display  

end  

and go to step 

8. 

CAL key 

To cancel the new setting and 

go back to step 4. 

Quitting the operation 

8  The balance displays the next menu of the function 

table. Press the CAL key to exit the clock and 
calendar function and return to the weighing mode. 

Notes 
Do not enter invalid values such as a non-existing date when setting the time and date. 

When the clock backup battery has been depleted, the balance displays

 

 

rtc

 

pf

 

. The dead 

battery only affects the clock and calendar function. When this happens, press any key 
and set the time and date. The clock and calendar function will continue to work normally 
as long as the AC adapter is connected to the balance

.

 

 
 

 
 
 
 
 
 
 
 

To  the date

setting mode

Confirm

Summary of Contents for FX-1200CT

Page 1: ...FZ CT SERIES FZ 1200CT FZ 700CT FZ 500CT FX CT SERIES FX 1200CT FX 700CT FX 500CT Carat Balance INSTRUCTION MANUAL 1WMPD4002564 ...

Page 2: ...ted into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice Windows Word and Excel are the registered trademarks of the Microsoft Corporation ...

Page 3: ...ration only for the FZ CT series 14 7 3 Calibration Test Using the Internal Mass Only for the FZ CT series 15 7 4 Calibration Using an External Weight 16 7 5 Calibration Test Using an External Weight 18 7 6 Correcting the Internal Mass Value 1 Only for the FZ CT series 20 7 7 Correcting the Internal Mass Value 2 Only for the FZ CT series 21 8 FUNCTION SWITCH AND INITIALIZATION 22 8 1 Permit or Inh...

Page 4: ... 49 11 UNDERHOOK 53 12 PROGRAMMABLE UNIT 54 RS 232C SERIAL INTERFACE 13 RS 232C SERIAL INTERFACE 55 14 CONNECTION TO PERIPHERAL EQUIPMENT 56 14 1 Connection to the AD 8121B Printer 56 14 2 Connection to a Computer 57 14 3 Using Windows Communication Tools WinCT 57 15 COMMANDS 59 15 1 Command List 59 15 2 Acknowledge Code and Error Codes 60 15 3 Settings Related to RS 232C 61 MAINTENANCE 16 MAINTEN...

Page 5: ...ault settings of the unit of weight are the metric carat ct and grams g The unit can be switched by pressing the MODE key Standard RS 232C serial interface to communicate with a computer and to output the Good Laboratory Practice GLP Good Manufacturing Practice GMP compliant report Statistical calculation mode to statistically calculate the weight data and display or output the sum maximum minimum...

Page 6: ...interference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in the U S A Compliance with EMC Directives This device features radio interference suppression and safety regulation in compliance with the following Council Directives Council directive 2004 108 EC EN6...

Page 7: ...alance in the future See the illustration to confirm that everything is contained For accurate weighing install the breeze break before use The AC adapter plug is protected against dust and may be difficult to insert When inserting the plug turn the plug while pushing on it Note Please confirm that the AC adapter type is correct for your local voltage and receptacle ...

Page 8: ...e illustration on page 5 3 Adjust the leveling feet to level the balance Confirm it using the bubble spirit level 4 Confirm that the AC adapter type is correct for the local voltage and power receptacle type 5 Connect the AC adapter to the balance Warm up the balance for at least one hour with nothing on the weighing pan Note The AC adapter plug is protected against dust and may be difficult to in...

Page 9: ...which produces magnetic fields Install the balance in a stable place avoiding vibration and shock Corners of rooms on the first floor are best as they are less prone to vibration The weighing table should be solid and free from vibration drafts and as level as possible Level the balance by adjusting the leveling feet and confirm it using the bubble spirit level Make sure that the bubble is in the ...

Page 10: ...d the sample Make each weighing gently and quickly to avoid errors caused by the evaporation of moisture from the sample or the absorption of moisture by the sample Use a pair of long tweezers to place a sample on the pan to avoid placing a hand inside the breeze break Do not drop things upon the weighing pan or place a sample on the pan that is beyond the balance weighing capacity Place a sample ...

Page 11: ...ld detergent to clean The edge of the weighing pan is sharp Use caution when cleaning the pan Breeze break standard accessory An antistatic treatment has been applied to the breeze break components Use a soft dry lint free cloth to clean Cleaning with a cloth that is moistened with water or a mild detergent repetitively or washing with water may degrade the antistatic effect 3 4 Power Supply When ...

Page 12: ...eighing mode turns the minimum weighing value on or off In the counting or percent mode enters the sample storing mode Enters the function table mode Refer to 9 FUNCTION TABLE Switches the weighing units stored in the function table Refer to 9 9 Storing Units Enters the response adjustment mode Cancels the operation when performing settings With the FZ CT series enters the calibration mode using t...

Page 13: ...he container 3 Wait for the stabilization indicator to be displayed Read the value While the stabilization indicator is on pressing the PRINT key will output the weight value using the RS 232C serial interface Note Peripheral equipment that is sold separately such 4 Remove the sample and container from the pan Notes When other units of weight are stored as described in 9 9 Storing Units pressing t...

Page 14: ...en release the key 2 Press the MODE key to select a rate of the response adjustment Either FAST MID or SLOW can be selected 3 After a few seconds of inactivity or when the PRINT key is pressed the balance displays eend Then it returns to the weighing mode and displays the updated response indicator The response indicator remains displayed for a while Note Setting the response rate automatically ch...

Page 15: ... the first time The balance has been moved The ambient environment has changed Do not allow vibration or drafts to affect the balance during calibration To output the GLP GMP compliant report using the RS 232C serial interface set GLP output info of Data output dout For details refer to 9 FUNCTION TABLE Calibration test is available only when GLP output info of Data output dout is set to 1 or 2 Ca...

Page 16: ...ance displays end after calibration If the GLP output info parameter of the function table is set to 1 or 2 the balance displays glp and outputs Calibration Report using the RS 232C interface For details on the calibration report format refer to 10 2 GLP Report 5 The balance automatically returns to the weighing mode after calibration About the internal mass The value of the internal mass may chan...

Page 17: ...ut info to 1 or 2 3 Press and hold the CAL key until CC in is displayed then release the key 4 The balance measures the zero point Prevent vibration and drafts to affect the balance 5 The measured zero point data is di played 6 The balance measures the internal mass Prevent vibration and drafts to affect the balance 7 The value of the internal mass is displayed The normal range of the value is as ...

Page 18: ...e SAMPLE key and proceed to step 4 If you use the calibration weight value stored in the balance proceed to step 5 4 Specify the calibration weight value as follows SAMPLE key To switch the display condition to All of the segments blinking calibration weight selection mode or The last three digits blinking value adjustment mode RE ZERO key To increase the value MODE key To decrease the value To se...

Page 19: ...ion or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays 1 glp and outputs Calibration Report 9 The balance automatically returns to the weighing mode 10 Press the MODE key to select the gram mode Place the calibration weight on the pan and confirm that the value displayed is within 0 002 g of the calibration we...

Page 20: ...u want to change the target weight a list of usable weights is shown on page 13 press the SAMPLE key and proceed to step 4 If you use the target weight value stored in the balance proceed to step 5 4 Specify the target weight value as follows SAMPLE key To switch the display condition to All of the segments blinking target weight selection mode or The last three digits blinking value adjustment mo...

Page 21: ...arget weight value 6 Place the displayed target weight on the pan and press the PRINT key The balance measures the target weight and displays the measured value Do not allow vibration or drafts to affect the balance 7 The balance displays end Remove the weight from the pan 8 When the GLP output is set the balance displays glp1 and outputs Calibration Test Report 9 The balance automatically returns...

Page 22: ... 8 FUNCTION SWITCH AND INITIALIZATION and set the internal mass value correction switch to 1 3 Press the MODE key to select the gram mode 4 Press and hold the SAMPLE key to display ba5fnc 5 Press the SAMPLE key several times until C5 in 1 is displayed 6 Press the PRINT key Correct the internal mass value using the following keys RE ZERO key To increase the value by one MODE key To decrease the val...

Page 23: ...o 1 2 Press the MODE key to select the gram mode 3 Press and hold the SAMPLE key to display ba5fnc 4 Press the SAMPLE key several times until C5 in 21 is displayed If C5 in 2 is not displayed perform step 1 5 Press the PRINT key to display Cal 0 Refer to 7 4 Calibration Using an External Weight to perform the calibration 6 After calibration remove the weight The balance displays Close the door of ...

Page 24: ...ing these parameters Each switch can select either permit or inhibit The inhibit protects parameters against unintentional operations FZ CT series FX CT series Function switch Function table Calibration using the internal mass Calibration using an external weight Internal mass correction Function table Calibration using an external weight Switches FZ CT series Switches FX CT series ...

Page 25: ...key once again 8 2 Initializing the Balance This function returns the following parameters to factory settings Calibration data Function table The sample unit mass value counting mode 100 reference mass value percent mode External calibration weight Function switch settings Statistical data Internal mass correction value Only for the FZ CT series Note Be sure to calibrate the balance after initial...

Page 26: ...de A for Data output mode and 10 digits for Auto print difference 9 2 Display and Keys Display Key Description The symbol indicates that the parameter displayed is in effect When pressed and held in the weighing mode enters the function table mode Selects the class or item in the function table mode Changes the parameter When a class is displayed moves to an item in the class When an item is displ...

Page 27: ...Turns on the weighing mode display when the AC adapter is connected 0 OFF poff Auto display OFF 1 ON 10 minutes Turns off the display after 10 minutes of inactivity 0 Displays rng Display at start 1 Does not display Select whether or not to display the minimum weighing value at weighing start 0 Does not sound ba5fnc Environment Display beep Beep 1 Sounds Select whether or not to sound the beep whe...

Page 28: ...o print polarity 2 Both Regardless of displayed value 0 10 digits 1 100 digits ap b Auto print difference 2 1000 digits Difference between reference value and displayed value 0 Every display refresh 1 2 seconds 2 5 seconds 3 10 seconds 4 30 seconds 5 1 minute 6 2 minute 7 5 minute int Interval time 8 10 minute Interval time for the interval output mode With prt 6 0 No output 1 Time only 2 Date onl...

Page 29: ...d ID number Refer to 10 ID NUMBER AND GLP REPORT 0 Normal weighing mode 1 Capacity indicator apf Application function 2 Statistical calculation mode 0 Number of data sum 1 Number of data sum maximum minimum average range maximum minimum 2 Number of data sum maximum minimum average range maximum minimum standard deviation coefficient of variation Aap fnc Application 5taf Statistical function mode o...

Page 30: ...es slight fluctuation of a weight value Used to prevent a weight value from drifting due to vibration or drafts Note With Hold function Hold set to ON 1 this item is used to set the stabilization range Hold function Hold Animal weighing mode This function is used to weigh an unstable sample such as an animal When the weight data is over the weighing range from zero and the display fluctuation is w...

Page 31: ... pnt The decimal point format can be selected Auto display ON p on When the AC adapter is connected the display is automatically turned on in the weighing mode without the ON OFF key operation Used when the balance is built into an automated system A one hour warm up is necessary for accurate weighing Auto display OFF poff When the weight value is 0 5 g or less and no operation such as key input o...

Page 32: ...Example For monitoring data on the AD 8920A remote display Required setting dout prt 3 Stream mode ba5fnc 5pd Display refresh rate 5if bp5 Baud rate Caution The balance may not transmit the data completely at the specified display refresh rate and baud rate Set the baud rate higher Key mode B When the PRINT key is pressed the balance outputs the weight data regardless of the display condition The ...

Page 33: ... data Output example Print example DP Dump print format 5if type 1 This format is suitable for the peripheral equipment that prints the received data as is When an AD 8121B printer is used set the printer to MODE 3 This format consists of sixteen characters excluding the terminator A header of two characters indicates the balance condition No overload header is used The polarity sign is placed bef...

Page 34: ...e balance condition The polarity sign is used only for negative data The weight data uses spaces in place of the leading zeros The character length of this format changes dependent upon the unit NU numerical format 5if type 4 This format outputs only numerical data This format consists of nine characters excluding the terminator The polarity sign is placed before the data with the leading zeros If...

Page 35: ...33 9 7 Data Format Examples LF ...

Page 36: ...34 ...

Page 37: ...ed and installed at the factory For details about the units and modes see the table below Name unit mode Display Function table Storing mode Conversion factor 1 g Gram g g 1 g Counting mode Percent mode Ounce Avoir 28 349523125 g Pound 453 59237 g Pound Ounce 1 lb 16 oz 1 oz 28 349523125 g Troy Ounce 31 1034768 g Metric Carat 0 2 g Momme 3 75 g Pennyweight 1 55517384 g Grain UK 0 06479891 g Tael H...

Page 38: ...Pound 0 555 0 335 0 224 0 000005 Pound Ounce 0L 8 88 oz 0L 5 36 oz 0L 3 60 oz 0L 0 01 oz Troy Ounce 8 10 4 88 3 27 0 00005 Metric Carat 1260 760 510 0 001 Momme 67 2 40 5 27 2 0 001 Pennyweight 162 0 97 7 65 5 0 001 Grain UK 3888 2345 1574 0 02 Tael HK general Singapore 6 66 4 02 2 69 0 00005 Tael HK jewelry 6 73 4 06 2 72 0 00005 Tael Taiwan 6 72 4 05 2 72 0 00005 Tael China 8 06 4 86 3 26 0 0000...

Page 39: ...Specify a unit or mode in the order to be displayed using the following keys SAMPLE key To sequentially display the units RE ZERO key To specify a unit or mode The stabilization indicator appears when the displayed unit or mode is specified 5 Press the PRINT key to store the units or modes The balance displays end and then displays the next menu of the function table 6 Press the CAL key to exit th...

Page 40: ...t displayed store the counting mode as described in 9 9 Storing Units Storing a sample unit mass 2 Press the SAMPLE key to enter the sample unit mass storing mode Even in the storing mode pressing the MODE key will switch to the next mode 3 To select the number of samples press the SAMPLE key several times It may be set to 5 10 25 50 or 100 Note A greater number of samples will yield more accurate...

Page 41: ...d on the pan While the stabilization indicator is on pressing the PRINT key will output the weight value number of objects using the RS 232C serial interface Note Peripheral equipment that is sold separately such as a printer or a personal computer is required Counting mode using the ACAI function The ACAI is a function that improves the accuracy of the unit mass automatically by increasing the nu...

Page 42: ...e the sample to be set as the 100 reference mass on the pan or in the container 5 Press the PRINT key to store the reference mass The balance displays 100 00 The decimal point position depends on the reference value The reference mass stored even if the AC adapter is removed is maintained in non volatile memory Note If the balance judges that the mass of the sample is too light to be used as a ref...

Page 43: ...mode the indicator displays the weight data relative to the weighing capacity in percentage Zero 0 maximum capacity 100 Statistical calculation mode apf 2 The mode statistically calculates the weight data and displays or outputs the results For details refer to 9 13 Statistical Calculation Mode ...

Page 44: ...oefficient of variation and relative error are obtained by the equation below Standard deviation Standard deviation Coefficient of variation CV Average Maximum value Average Relative error of maximum value Average Minimum value Average Relative error of minimum value Average 9 13 1 Getting Started Switching to the Statistical Function Mode Changing The Function Table 1 Press and hold the SAMPLE ke...

Page 45: ...ter the data is entered In this case clear the data as described on page 45 and select the unit using the MODE key When the unit used for the statistical calculation mode is to be enabled upon power on select the unit in Unit Unit of the function table beforehand 9 13 2 Using The Statistical Calculation Mode Entering data for statistical calculation Use the following keys to operate the statistica...

Page 46: ... in Statistical function mode output items 5taf and Notes When the number of data is 1 the coefficient of variation and the relative error are displayed as When the average is 0 the coefficient of variation and the relative error are as Statistical items are indicated on the upper left of the display using the following symbols Symbol Statistical item 5um Sum 1max Maximum 1min Minimum 1r Range Max...

Page 47: ...lete the latest data and exclude it from statistical calculation The number of data decreases by 1 when the balance returns to the weighing mode Clearing the statistical data All the statistical data will be deleted and the number of data will be 0 zero 1 In the weighing mode press the MODE key several times to display 2 Press the PRINT key to display 3 Press the RE ZERO key to display 4 Press the...

Page 48: ...mode Entering the upper limit value 6 With displayed press the PRINT key The current setting of the upper limit value is displayed with all the digits blinking When the current setting is not to be changed press the PRINT or CAL key to proceed to step 7 When the current setting is to be changed press the RE ZERO key Change the setting using the following keys SAMPLE key To select the digit to chan...

Page 49: ...th all the digits blinking When the date is correct and the operation is to be finished press the CAL key and go to step 8 When the date is correct and the time is to be confirmed press the SAMPLE key and go to step 6 When the date is not correct and is to be changed press the RE ZERO key and go to step 5 Note The year is expressed using a two digit format For example the year 2012 is expressed as...

Page 50: ...e the value by one MODE key To decrease the value by one PRINT key To store the new setting display end and go to step 8 CAL key To cancel the new setting and go back to step 4 Quitting the operation 8 The balance displays the next menu of the function table Press the CAL key to exit the clock and calendar function and return to the weighing mode Notes Do not enter invalid values such as a non exi...

Page 51: ...n test using an external weight Title block and End block for the weight data 10 1 Setting the ID Number 1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed 2 Press the SAMPLE key several times to display id 3 Press the PRINT key Set the ID number using the following keys SAMPLE key To select the digit to change the value The selected digit blinks RE ZERO key MODE key T...

Page 52: ...y for the FZ CT series When the setting is info 1 When the setting is info 2 Calibration test report using the internal mass Only for the FZ CT series Calibration test does not perform calibration When the setting is info 1 When the setting is info 2 ...

Page 53: ...tting is info 2 Calibration test report using an external weight Calibration test does not perform calibration When the setting is info 1 When the setting is info 2 With FZ CT series the time and date are added to output data With FZ CT series the time and date are added to output data ...

Page 54: ...r use MODE 3 of the AD 8121B printer Operation 1 With the weight data displayed press and hold the PRINT key until S5tart is displayed The Title block is output 2 The weight data is output according to the parameter setting of the data output mode 3 Press and hold the PRINT key until recend is displayed The End block is output When the setting is info 1 When the setting is info 2 With FZ CT series...

Page 55: ...uilt in underhook is revealed by removing the plastic cap on the bottom of the balance Use the underhook as shown below Caution Do not apply excessive force to the underhook When not in use attach the plastic cap to prevent dust from getting into the balance Underhook Cap Bottom of the balance ...

Page 56: ... coefficient 4 The current coefficient is displayed with the first digit blinking When it is not to be changed press the CAL key and proceed to step 6 When it is to be changed press the RE ZERO key and proceed to step 5 Setting the coefficient 5 Set the coefficient using the following keys SAMPLE key To select a digit to change the value The selected digit blinks RE ZERO key To change the value MO...

Page 57: ...600 1200 2400 4800 9600 19200 bps Data bits 7 or 8 bits Parity Even Odd Data bits 7 bits None Data bits 8 bits Stop bit 1 bit Code ASCII D Sub 9 pin assignments Pin No Signal name Direction Description 1 No connection 2 TXD Output Transmit data 3 RXD Input Receive data 4 No connection 5 SG Signal ground 6 DSR Output Data set ready 7 RTS Input Request to send 8 CTS Output Clear to send 9 No connect...

Page 58: ...ut pause 0 0 0 0 1 2 bp5 Baud rate 2 2 2 2 btpr Data bit parity bit 0 0 0 0 Crlf Terminator 0 0 0 0 5if Serial interface type Data format 0 0 0 1 1 Set appropriate parameters for ap p Auto print polarity and ap b Auto print difference when auto print mode A or B prt 1 or 2 is selected Set the AD 8121B DIP switch No 3 to ON when unstable data is printed with prt 4 2 Set 1 when multiple lines are pr...

Page 59: ...nsmit the weight data to the personal computer The WinCT has three communication methods RsCom RsKey and RsWeight RsCom Can transmit commands to control the balance Can make bi directional communication between the balance and a personal computer using the RS 232C serial interface Can display or store the data using a text file format Can also print the data using a printer connected to the person...

Page 60: ...h 2 Controlling the balance using commands from a personal computer By using RsCom the personal computer sends commands such as re zero or send weight data to the balance and controls the balance 3 Printing the balance GLP report using your printer The balance GLP report can be printed using a printer connected to the personal computer 4 Receiving weight data at a certain interval The weight data ...

Page 61: ...ommands the S and E SCP commands behave the same Commands to control the balance CAL Same as the CAL key EXC Calibration using an external weight OFF Turns the display off ON Turns the display on P Same as the ON OFF key PRT Same as the PRINT key R Same as the RE ZERO key SMP Same as the SAMPLE key T Tare key Z Same as the RE ZERO key E SCT Same as the RE ZERO key U Same as the MODE key ID Request...

Page 62: ...command to control the balance and can not process it the balance transmits an error code EC Exx When the balance receives a command to control the balance and can process it the balance transmits the acknowledge code Among commands to control the balance the following transmit the acknowledge code both when the balance receives the command and when the balance has accomplished the command If the ...

Page 63: ...l noise or a parity error has occurred due to transmission error the balance transmits an error code In this case send the command again 15 3 Settings Related to RS 232C Concerning the RS 232C the balance has two functions Data output dout and Serial interface 5if Set each function as necessary ...

Page 64: ...Use caution when cleaning the pan Breeze break standard accessory An antistatic treatment has been applied to the breeze break components Use a soft dry lint free cloth to clean Cleaning with a cloth that is moistened with water or a mild detergent repetitively or washing with water may degrade the antistatic effect Do not disassemble the balance Contact the local A D dealer if the balance needs s...

Page 65: ... value Checking that the operating environment or weighing method is proper Operating environment Is the weighing table solid enough Is the balance level Refer to 3 1 Before Use Is the operating environment free from vibration and drafts Is there a strong electrical or magnetic noise source such as a motor near the balance Weighing method Is the weighing pan installed correctly Is the RE ZERO key ...

Page 66: ...ht error The calibration weight is too heavy Confirm that the weighing pan is properly installed Confirm the calibration weight value Press the CAL key to return to the weighing mode EC E21 Calibration weight error The calibration weight is too light Confirm that the weighing pan is properly installed Confirm the calibration weight value Press the CAL key to return to the weighing mode Overload er...

Page 67: ...firm the format baud rate and parity EC E01 Undefined command error An undefined command was received Confirm the command EC E02 Not ready A received command can not be processed e g The balance received a Q command but not in the weighing mode e g The balance received a Q command while processing a RE ZERO command Adjust the delay time to transmit a command EC E03 Timeout error If the timeout par...

Page 68: ...vice or repair contact the local A D dealer The balance is a precision instrument Use caution when handling the balance and observe the following when transporting the balance Use the original packing material Remove the weighing pan from the main unit ...

Page 69: ...rt to a personal computer FXi 08 Ethernet interface Used to connect the balance to a LAN The WinCT Plus data communication software is provided as an accessory and can perform the following Acquire data from multiple balances connected to a LAN Control these balances with commands Acquire data transmitted from balances Example When pressing the PRINT key of the balance data is output and is acquir...

Page 70: ... if the personal computer has no COM ports serial communication software such as WinCT can be used via USB connection AD 8920A Remote display Connected to the FZ CT and FX CT series balances using the RS 232C serial interface to display the weight data away from the balance or at a desirable viewing angle AD 8922A Remote controller Connected to the FZ CT and FX CT series balances using the RS 232C...

Page 71: ...he static charge on the sample tare or peripheral equipment and displays the result If those are found to be charged discharge them using the AD 1683 DC static eliminator AD 1687 Weighing environment logger A data logger equipped with 4 sensors for temperature humidity barometric pressure and vibration that can measure and store environmental data When connected to the RS 232C interface of the bal...

Page 72: ...nit mass 0 001 g Counting mode 3 Number of samples 5 10 25 50 or 100 pieces Minimum 100 reference mass 0 100 g Percent mode 3 Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 250 g 50 g 200 g 20 g 100 g 150 g 50 g 100 g 20 g 100 g 20 g 50 g Weighing pan diameter 90 mm Net weight Approx 3 8 kg External dimensions 198 W x 294 D x 315 H mm AC adapter Con...

Page 73: ...s 0 001 g Counting mode 3 Number of samples 5 10 25 50 or 100 pieces Minimum 100 reference mass 0 100 g Percent mode 3 Minimum 100 display 0 01 0 1 1 Depends on the reference mass stored External calibration weight 250 g 50 g 200 g 20 g 100 g 150 g 50 g 100 g 20 g 100 g 20 g 50 g Weighing pan diameter 90 mm Net weight Approx 3 4 kg External dimensions 198 W x 294 D x 315 H mm AC adapter Confirm th...

Page 74: ...72 20 EXTERNAL DIMENSIONS FZ 1200CT 700CT 500CT FX 1200CT 700CT 500CT Unit mm ...

Page 75: ...included in the weight data Mode Balance operational function Re zero To set the display to zero GLP Good Laboratory Practice GMP Good Manufacturing Practice Repeatability Variation in measured values obtained when the same weight is placed and removed repetitively Usually expressed as a standard deviation e g Standard deviation 1 digit This means that measured values fall within 1 digit in the fr...

Page 76: ... difference 26 apf Application function 27 42 ap p Auto print polarity 26 56 Application 27 42 Application function 27 42 ar d Zero after output 26 at f Auto feed 26 Auto display OFF 25 29 Auto display ON 25 29 Auto print difference 26 Auto print mode A 26 30 Auto print mode B 26 30 Auto print polarity 26 ave Average 44 Average 27 42 44 AX CARAT PAN B 68 AX CARAT PAN S 68 AX FXi 31 68 AX USB 9P EX...

Page 77: ...lf Terminator 27 56 CSV format 27 32 CvV Coefficient of variation 44 D Data bits 55 Data format 27 Data logger 69 Data output mode 26 Data output pause 26 DC static eliminator 69 DCE 55 57 Decimal point 25 29 Digit 25 29 73 Display at start 25 Display refresh rate 25 29 dout Data output 26 DP format 27 31 DTE 55 E e 64 e 64 EC E00 65 EC E01 65 EC E02 65 EC E03 65 EC E04 65 EC E06 65 EC E07 65 EC E...

Page 78: ... 42 44 Minimum 100 reference mass 70 71 Minimum unit mass 70 71 Minimum weighing value 70 71 MODE key 3 10 37 MT format 27 32 N NU format 27 32 Number of data 27 42 O One touch calibration 14 ON OFF key 10 23 Operating environment 70 71 P Pan support 5 Parity 55 38 40 Percent Mode 40 Permit 22 pnt Decimal point 25 29 poff Auto display OFF 25 29 p on Auto display ON 25 29 PRINT key 10 37 Printer 56...

Page 79: ... 43 Statistical function mode output items 27 5t b Stability band width 25 28 Stop bit 55 Stream mode 26 30 5um Sum 44 Sum 27 42 44 T Tabletop breeze break 69 Tare 11 38 40 Terminator 27 Timeout 27 Title block 52 Transmission form 55 Transmission rate 55 Transmission system 55 trc Zero tracking 25 t Up Timeout 27 type Data format 27 56 U Underhook 53 Units 10 35 37 Upper limit value 46 USB interfa...

Page 80: ...78 MEMO ...

Page 81: ...79 MEMO ...

Page 82: ...80 MEMO ...

Page 83: ......

Page 84: ...nsburg Deutschland Telefon 49 0 4102 459230 Telefax 49 0 4102 459231 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 대한민국 서울시 영등포구 여의도동 36 2 맨하탄 빌딩 8층 8th Floor Manhattan Bldg 36 2 Yoido dong Youngdeungpo ku Seoul KOREA 전화 82 2 780 4101 팩스 82 2 782 4280 A D RUS CO LTD Компания ЭЙ энд ДИ РУС 1213...

Reviews: