10. Maintenance
Refer to “
5. Installation and Precautions
” regarding use.
Refer to “
10.3. Error Code Table
” and corresponding mode for displayed error code.
Refer to “
7.
Sensitivity Adjustment
” regarding precision weighing.
Periodically check the accuracy of weighing. Adjust the scale (indicator), if it is moved to another location
or the environment has changed.
10.1. Repair
Do not disassemble / assemble the indicator without an authorized service engineer. Doing so may
cause an electric shock or damage to the indicator, etc. In this case, repair is not covered under warranty.
Contact your local A&D dealer if your indicator needs service or repair.
10.2. Error Display
In this situation
Confirm these items
Nothing displayed.
Scale does not turn on.
Is the main power the correct voltage?
Is the main power cord properly connected?
The scale does not
display zero at first.
Check around the weighing pan.
Is there anything on the weighing pan?
Refer to “
” and perform zero point
sensitivity adjustment.
is displayed.
Weighing error that meaning "Overloaded".
is displayed.
Indicates that the sample weight is too light to set the unit weight in
the counting mode.
is displayed.
Sensitivity adjustment error that meaning "Too light".
is displayed.
Displayed when the main power drops.
is displayed
and does not proceed.
The weighed value is unstable due to drift, vibration or other factors.
A breeze or vibration may be affecting measurement.
Check around the weighing pan.
Check the connection of the load cell cable.
No zero display when the display is turned on. Remove anything that
is on the pan.
Perform zero point sensitivity adjustment.
Fixed display.
Did you use the "hold function" ? Refer to the [FG-CWP Series
Instruction Manual] on the A&D website (
) and
change the function
.
Turn off scale and turn it on again.