8.2. Setting Units for Use
1. Refer to “
7.1. Sensitivity Adjustment Items
”, to enter the sensitivity
adjustment mode. The indicator displays
.
2. Press the MODE key three times to display
.
3. Press the TARE key to display the setting item.
4. To proceed without storing this setting item, press the ZERO key.
After displaying
, proceed to the unit setting mode.
(Display example:
pcs )
5. Select the unit to be used, using the following keys.
MODE Select the unit (Shift the blinking unit to the right)
PRINT Set whether the selected unit is available (0) or
not available (1). (Change the number of the
blinking digit to “0” or “1”.)
About the unit setting mode
Each of the display digits are assigned to a unit setting.
(The setting at the first and fourth digit is not available.)
Display
assignment
7
6
5
4
3
2
1
Unit
pcs
g
kg
Set
disabled oz
lb
t
The unit during selection is indicated.
With each digit, if setting to “0”, its unit is not available,
and if setting to “1”, its unit is available.
(Only when set to “1”, is a unit available.)
Even if a unit is stored, availability for use depends on
the weighing capacity and resolution.
Example) When setting to
pcs ”, only “pcs”
and ”kg” are available as units.
6. Press the TARE key to store the setting. After displaying
,
the indicator displays
, and the setting is complete.
If you want to proceed without storing the setting of
the unit setting mode, press the ZERO key.
After displaying
, the setting is complete.
Press
3 times
Press
Press
Press
(Select)
Press