9.
CALIBRATION
Calibration of the scale is required when it is initially installed, if it is moved often, or
it is moved a substantial distance. Calibration is also necessary in regular scale
maintenance due to normal mechanical wear-and-tear, changes in seasonal
temperature, humidity, air pressure, etc.
The scale is equipped with gravity compensation, which allows it to be calibrated in
one location and then adjusted to match the gravity acceleration at another location
where it will be used. But don’t worry about this, as far as you calibrate the scale
using a calibration weight and use it at same place.
The scale must perform "warm up" (with the AC adapter connected) for at least 30
minutes before starting calibration.
9-1.
Calibration Procedure Using a Weight
The scale should have power connected at least one-half hour to warm it up before
starting the calibration procedure.
1. Remove the calibration switch cover,
and press the calibration (CAL) switch.
The scale shows “
Cal
” in the count
display.
Press the CAL switch to exit without calibrating the scale.
2. Press the
ZERO
key to enter into zero &
span calibration mode.
The display flashes the required calibration
weight value.
If you know the exact weight value, or if you
wish to use a different weight,
use the
0
Æ
9
and
.
10-key pad to
.
88
Cal
10
88
1
pcs
[ Z E R O ] C a l i b r a t i o n
1234567890123
[ T A R E ] G r a v i t y A c c e l e r a t i o n
1234567890123
Calibration
switch cover
Calibration (CAL) switch
.
88
Cal
10
88
1
pcs
C a l i b r a t i o n w e i g h t ( k g )
Cal4567890123aaaaaaaaa
10.000
[ T A R E ] G r a v i t y A c c e l e r a t i o n
.
88
Cal
10
88
1
pcs
C a l i b r a t i o n w e i g h t ( k g )
Cal4567890123aaaaaaaaa
10.002
[ T A R E ] G r a v i t y A c c e l e r a t i o n
1234567890123
TUV
DEF
ABC
MNO
6
1
7
8
PQRS
2
3
JKL
GHI
0
9
WXYZ
4
#
5
C
.
31
Summary of Contents for FC-1000i
Page 60: ...Comparator upper limit 12345 Codabar NW 7 A17 000012345A 58 ...
Page 79: ...World map 77 ...
Page 82: ...MEMO 80 ...
Page 83: ...MEMO 81 ...
Page 84: ... key Space code Standby indicator 82 ...
Page 85: ......
Page 86: ......