28
Change the function table
Selecting a way to set the density of a liquid
Select the liquid density input method from the function table below. The function table
is available only when the density measurement mode is selected.
Class Item
Param-
eter
Description
0
Water temperature
func
ldin
Liquid density input
1
Liquid density
The way to input
liquid density.
Factory setting
Setting the density of a liquid
1. Press the UNITS key to select SG .
2. Press and hold the UNITS key to display liquid density input mode.
ldin
=
0
: Water temperature
This shows 25 °C.
ldin
=
1
: Liquid density
This shows
ρ
= 1.0000 (g/cm
3
).
3. Using the RE-ZERO (to increment the value) and SAMPLE key (to shift the
selected digit), set the value and press the PRINT key to store.
To cancel the setting procedure and return to the density measuring mode, press
the
UNITS key. The input value is not stored.
The relation between the water temperature and density is shown below.
°C
+0 +1 +2 +3 +4 +5 +6 +7 +8 +9
0
0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
10
0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20
0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30
0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40
0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
50
0.98804 0.98758 0.98712 0.98665 0.98618 0.98570 0.98521 0.98471 0.98422 0.98371
60
0.98320 0.98268 0.98216 0.98163 0.98110 0.98055 0.98001 0.97946 0.97890 0.97834
70
0.97777 0.97720 0.97662 0.97603 0.97544 0.97485 0.97425 0.97364 0.97303 0.97242
80
0.97180 0.97117 0.97054 0.96991 0.96927 0.96862 0.96797 0.96731 0.96665 0.96600
90
0.96532 0.96465 0.96397 0.96328 0.96259 0.96190 0.96120 0.96050 0.95979 0.95906
Example of density measurement
Selecting the SG measurement mode
1. Press the UNITS key to select SG .
(The weight unit is “g”.)
The weighing unit is “g”.
The display shows that
HI
blinks and the balance
is measuring weight in air.
When the display does not show zero, press the RE-ZERO key to set the
display to zero.
t
---
25
SG
d1.0000
SG
SG indicator
0.00
SG
HI
SG indicator
0.00
SG
HI
Measuring
weight in air.