Error code
Probable cause
Correction
E55
、
E56
、
E57
Exhaust
error
Relax and do not move during the
measurement. If this error occurs frequently,
repair is necessary.
E60
Interval setting error
Enter the interval parameters correctly.
E70
、
E71
、
E72
、
E73
RS-232C error
Reconnect the communication cable. If the
error persists, repair is necessary.
E74
Voltage decreases during
communication.
Replace the batteries with new ones and
restart communication.
E75
Protocol error due to
external equipment
Reconnect the communication cable. If the
error persists, repair is necessary.
14-1-2 Guidance messages
The table below lists the guidance messages that may appear in the Receive Data/Send Setting
Data windows.
y
Messages when data is received:
Message
Correction
Failed to initialize RS233C port
Automatic recognition of ABPM failed.
Failed to receive data
Failed to read measurement settings
Failed to read measurement data
Failed to erase measurement data
Failed to setup the clock
y
Confirm that the communication cable is
connected to the blood pressure monitor.
y
Confirm that the USB driver is functioning
correctly.
y
Disconnect all the cables. Restart Doctor
Pro3.
y
Messages when data is sent:
Message
Correction
Failed to initialize RS233C port
Automatic recognition of ABPM failed.
Failed to send data
Failed to erase measurement data
Failed to erase measurement settings
Failed to set measurement parameters.
Failed to setup the clock
y
Confirm that the communication cable is
connected to the blood pressure monitor.
y
Confirm that the USB driver is functioning
correctly.
y
Disconnect all the cables. Restart Doctor
Pro3.
The device connected differs from the one at
the time of automatic recognition.
y
Use the blood pressure monitor that was
automatically recognized.
y
Click [Cancel] or restart Doctor Pro3.