77
4. Weighing pan check
Make sure that there is nothing on the weighing pan.
1
Skip button
2
Back button
3
Confirm button
Name
Description
1
Skip button
Skips this check item and proceeds to the next item.
2
Back button
Returns to the previous screen.
3
Confirm button
Considers this check item as OK and proceeds to the next item.
5. Calibration Test (Before Sensitivity Adjustment)
The calibration test is performed before sensitivity adjustment.
When the test is completed, the [Calibration test result] screen is displayed.
When the internal weight is used, this is automatically performed. When an external weight is used,
follow the instruction.
Do not apply vibration and the like to the balance during the calibration test.
3
1
4
2
5
6
Name
Setting value
(setting range)
Description
1
Weight to be
used
Internal weight ,
External weight
Select the weight to be used.
2
Back button
-
Cancels and returns to the previous screen.
3
Result display
-
Displays the calibration test result.
4
Skip button
-
Skips this check item and proceeds to the next item.
5
Execute button
-
Executes the calibration test.