When a command controlling the balance is sent to the balance:
If the balance cannot execute the command, it sends an error code (EC, Exx).
If the balance can execute the command, it sends the AK code (acknowledgment, ASCII 06h).
Command
Description
ON
Turns the display on.
P
Turns the display on/off. (Only when the display is on.)
R, RZ
Same as the RE-ZERO button
.
T, TR
Same as the TARE button
.
ZR
Zero
If the load is within ±2% of the capacity from the initial zero point, the zero
point is updated, the tare value is cleared and the display is set to zero. If the
load exceeds ± 2%, no processing is done.
CAL
Sensitivity adjustment with the internal weight.
EXC
Sensitivity adjustment with an external weight.