background image

BLADE REMOVAL position

3.9a

3.9b

3.9c

3.9d

3.9e

3.9f

Charger 

Australia/China

International plug adapters

UK

Europe/

South America

North America/

Japan/China

4a

4b

en

16

17

en

Figure 3.9: Blade removal sequence

The blade is removed by turning the device upside down (

Figure 3.9a

), press the thumb 

firmly against the exposed heel of the Camera Module to avoid accidental detachment 

during blade removal (

Figure 3.9b

). Rotate and lock the Camera Module into the ‘BLADE 

REMOVAL’ position (

Figure 3.9c

). Using the thumb and index finger of the free hand pinch 

the ridged grip zones on the open end of the blade and squeeze firmly together (

Figure 

3.9d

). Whilst the grip is squeezed a force away from the Handle should be applied to clear 

the locking catch on the Camera Module (

Figure 3.9e

). The blade will slide off the Camera 

Module (

Figure 3.9f

) without making contact with the contaminated tip of the blade. It 

should then be disposed of in accordance with standard departmental practices.

The Video Viewer should be charged when the LED light indicator is flashing, indicating 

that there is insufficient charge to use the device correctly and safely. The Video Viewer is 

supplied with a charger and 4 international plug adapters (

Figure 4a

). The charging time 

for a completely uncharged battery is approximately 2 hours. It is safe to leave the Video 

Viewer fitted with the charger indefinitely if desired.
To begin charging, attach the appropriate plug adapter to the charger (

Figure 4b

) and 

connect to the mains power supply. Once connected, align the convex point of the Video 

Viewer connector with the concave point of the Video Viewer Charger connector to confirm 

correct charging direction (

Figure 4c

). Attachment is made easy by the magnetic pins in 

the connector of the Video Viewer and Video Viewer Charger (

Figure 4d

).

CAUTION:

 Care should be taken during charging operations and limit personnel contact 

with any electrical sources.

Figure 4a: Video Viewer Charger and international plug adapters

Figure 4b: Attach the appropriate plug to the charger

2.7 

Removing the blade after use

3. 

Battery and charging information

3.1 

Charging instructions for the Video Viewer

CAUTION:

 

The blade is designed with a fail-safe feature and can only be removed when 

the Camera Module is in the BLADE REMOVAL position.

The APA

 Stylet is designed to work with the APA

 U-DAB by providing an angled shape to 

facilitate ETT placements during laryngoscopy. It can be used with ETT up to size 9.0mm 

(I.D) or 12.8mm (O.D).
Railroad the stylet through the appropriate ETT, and after obtaining a view of the laryngeal 

inlet, advance the ETT into the trachea. The APA

 stylet is reusable, and should be 

cleaned and disinfected before first use and between each patient use similar to the video 

laryngoscope, refer to Section 4.

CAUTION:

 Do not allow the stylet to protrude beyond the end of the ETT during intubation.

2.6.5 APA

 Stylet

Summary of Contents for APA Video Laryngoscope

Page 1: ...it ISTRUZIONI PER L USO ja 使用説明書 ko 사용설명서 nl GEBRUIKSAANWIJZING no BRUKSANVISNING pl INSTRUKCJA UŻYCIA pt INSTRUÇÕES DE UTILIZAÇÃO sv BRUKSANVISNING tr KULLANMA TALİMATI zh 使用说明 Venner Medical Singapore Pte Ltd 35 Joo Koon Circle Singapore 629110 www vennermedical com Advena Ltd Tower Business Centre 2nd Flr Tower Street Swatar BKR 4013 Malta www advenamedical com ...

Page 2: ...lade 15 2 6 5 APA Stylet 16 2 7 Removing the blade after use 16 3 Battery and charging information 17 3 1 Charging instructions for the Video Viewer 17 3 2 Video Viewer LED light indicators 18 3 3 Battery replacement in the Handle 19 4 Decontamination 20 4 1 Preparation 20 4 2 Cleaning 20 4 3 Disinfection 21 4 4 Sterilisation 21 4 4 1 Low temperature 21 4 4 2 Steam sterilisation 21 4 5 APA IP Shie...

Page 3: ...ryngoscopy in routine and difficult airway intubations It may be used to facilitate endotracheal intubation as part of general anaesthesia or CPR cardiopulmonary resuscitation during cardiac arrest It is suitable for use in pre hospital hospital and extended care facilities The video laryngoscope is reusable DEHP free and latex free It is fully detachable into three components APA Video Viewer APA...

Page 4: ...ble To record video the video cable may be connected to a suitable recording device The Video Viewer cannot be autoclaved and must be cleaned and disinfected between first use and each patient use refer to Section 4 The Handle incorporates and is powered by a single replaceable alkaline or lithium AA 1 5V battery and the Camera Module incorporates a high intensity light source and a miniature came...

Page 5: ...n a non sterile reusable state It will be necessary to clean it before use refer to Section 4 Check the battery of the laryngoscope Before using the device as a standard laryngoscope ensure that the Camera Module illuminates when the device locks in the ON position CAUTION The AA 1 5V battery MUST be changed regularly Inadequate battery level may cause the laryngoscope to fail suddenly without war...

Page 6: ...place until the limit of travel is reached Figure 3 3 To confirm correct attachment rotate the Camera Module into the ON position the Video Viewer screen should then turn on If using as a video laryngoscope the Video Viewer should be attached to the laryngoscope before using on a patient CAUTION If an image cannot be seen on the Video Viewer check that it is correctly attached to the Handle Avoid ...

Page 7: ... the epiglottis is elevated by lifting the Handle of the laryngoscope exposing the laryngeal inlet With direct visualisation of the laryngeal inlet a correctly sized ETT EndoTracheal Tube can be introduced into the trachea Ensure that the ETT is adequately lubricated prior to use An introducer stylet may be used to assist the introduction Correct placement of the ETT should be both visual and by t...

Page 8: ...t With direct visualisation of the laryngeal inlet a correctly sized ETT EndoTracheal Tube can be introduced into the trachea Ensure that the ETT is adequately lubricated prior to use An introducer stylet may be used to assist the introduction Correct placement of the ETT should be both visual and by the use of capnography Figure 3 7 Introducing the device with MIL Blade Figure 3 8 Connecting the ...

Page 9: ... fitted with the charger indefinitely if desired To begin charging attach the appropriate plug adapter to the charger Figure 4b and connect to the mains power supply Once connected align the convex point of the Video Viewer connector with the concave point of the Video Viewer Charger connector to confirm correct charging direction Figure 4c Attachment is made easy by the magnetic pins in the conne...

Page 10: ...the battery level is sufficient If the LED light indicator starts to flash during a procedure it means that there is limited battery The procedure may continue at the clinician s discretion The Camera Module is supplied with an alkaline AA 1 5V battery When used as a standard laryngoscope the Camera Module illuminates when the device locks in the ON position The battery should be replaced regularl...

Page 11: ...sed bag for transportation to the sterile processing unit for sterilisation CAUTION Only the Handle and Camera Modules may be sent for reprocessing Prepare or dilute a suitable disinfectant in accordance with the manufacturer s instructions Visit the U S Environmental Protection Agency EPA website for a list of registered disinfectants Spray or wipe the components with a suitable disinfectant i e ...

Page 12: ...battery If the illumination still does not work contact the local distributor for servicing refer to Section 5 5 Service and maintenance 6 Troubleshooting 6 1 As a video laryngoscope 6 2 As a laryngoscope The video laryngoscope should be inspected at regular intervals for Visible signs of external damage Scratches or defects on the Camera Module lens Build up of material on connection pin area of ...

Page 13: ...IN LIEU OF ALL OTHER WARRANTIES WRITTEN OR ORAL EXPRESS OR IMPLIED STATUTORY OR OTHERWISE NO IMPLIED STATUTORY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY AAM HEALTHCARE SHALL NOT BE LIABLE FOR ANY DAMAGES SUSTAINED BY BUYER ARISING FROM DELAY IN THE REPLACEMENT OR REPAIR OF PRODUCTS UNDER THE ABOVE WARRANTY Terms and conditions may change without prior notice Dif...

Page 14: ...g instructions for this product Residents outside the European Union must dispose or recycle this product in accordance with local laws or regulations that apply Device should be stored and used between 4ºC and 40ºC Keep dry Keep away from sunlight Date of manufacture Legal manufacturer EU and Manufactured for US Authorised representative in the European Community European Conformity mark Class 1 ...

Page 15: ...thcare All rights reserved AAM Healthcare APA and Venner are trademarks of the Venner Group of Companies This product is covered by patents and patent applications of the Venner Group of Companies in the US and other territories Visit www AAMHealthcare com intellectual property 2266 H January 2019 ...