To use the device as a laryngoscope, align the notch on the Camera Module with the 

steel pin on the Handle and attach them together (

Figure 2

). An audible “click” may 

be heard when the Camera Module is fitted correctly. Ensure that the Camera Module 

rotates smoothly and locks in the ‘OFF’, ‘BLADE REMOVAL’ and ‘ON’ positions (

Figure 2.1

). 

Check that the laryngoscope powers up in the ‘ON’ position. A bright light should 

appear at the tip of the Camera Module. To preserve battery, store in the ‘OFF’ position.

CAUTION:

 

Do not look directly into the light source of the Camera Module.

Figure 2: Assembling the laryngoscope 

Figure 2.1: Lock positions

To use the device as a video laryngoscope, first rotate the connector of the Video 

Viewer outwards (

Figure 3

). With the Handle in one hand, fold back the connector cap 

on its flexible spine from the top of the Handle (

Figure 3.1

). With the other hand,hold the 

Video Viewer firmly and ensure that the convex point on the Video Viewer matches 

the concave point on the Handle (

Figure 3.2

). The fold back connector cap should 

be aligned with the backside of the Video Viewer. Proceed to slot the Video Viewer 

connector into the Handle connector recess, firmly pressing downwards into place, 

until the limit of travel is reached (

Figure 3.3

). To confirm correct attachment rotate 

the Camera Module into the ‘ON’ position, the Video Viewer screen should then turn 

on. If using as a video laryngoscope, the Video Viewer should be attached to the 

laryngoscope before using on a patient.

CAUTION:

 

If an image cannot be seen on the Video Viewer, check that it is correctly 

attached to the Handle.

Avoid applying any pressure on the Video Viewer screen while attaching the devices. 

Take care not to dislodge the connection between the devices at any time.

Figure 3: Rotate the connector outwards

Figure 3.1: Fold back the connector cap

Figure 3.2: Ensure correct direction by matching the convex and concave points

Figure 3.3: Attach devices together

To fit a new blade the Camera Module needs to lock in the ‘ON’ position. Open the 

blade packaging pouch and without fully removing the blade from the packaging, 

slide it onto the Camera Module (

Figure 3.4

). An audible “click” may be heard when 

the blade is fitted correctly (In the case of the Miller, matching an identical colour 

code between the blade and Camera Module suggests a correct pairing). Proceed 

to remove the blade packaging completely without touching the blade. 

CAUTION:

 

Do not use the blade if the packaging has been opened or damaged. 

Take note to pair the same style of blade and Camera Module, and check that the 

blade is correctly fitted to the Camera Module prior to using on a patient.

2.4 

Assembling the device

2.5 

Fitting a new blade

Summary of Contents for APA Video Laryngoscope

Page 1: ...it ISTRUZIONI PER L USO ja 使用説明書 ko 사용설명서 nl GEBRUIKSAANWIJZING no BRUKSANVISNING pl INSTRUKCJA UŻYCIA pt INSTRUÇÕES DE UTILIZAÇÃO sv BRUKSANVISNING tr KULLANMA TALİMATI zh 使用说明 Venner Medical Singapore Pte Ltd 35 Joo Koon Circle Singapore 629110 www vennermedical com Advena Ltd Tower Business Centre 2nd Flr Tower Street Swatar BKR 4013 Malta www advenamedical com ...

Page 2: ...lade 15 2 6 5 APA Stylet 16 2 7 Removing the blade after use 16 3 Battery and charging information 17 3 1 Charging instructions for the Video Viewer 17 3 2 Video Viewer LED light indicators 18 3 3 Battery replacement in the Handle 19 4 Decontamination 20 4 1 Preparation 20 4 2 Cleaning 20 4 3 Disinfection 21 4 4 Sterilisation 21 4 4 1 Low temperature 21 4 4 2 Steam sterilisation 21 4 5 APA IP Shie...

Page 3: ...ryngoscopy in routine and difficult airway intubations It may be used to facilitate endotracheal intubation as part of general anaesthesia or CPR cardiopulmonary resuscitation during cardiac arrest It is suitable for use in pre hospital hospital and extended care facilities The video laryngoscope is reusable DEHP free and latex free It is fully detachable into three components APA Video Viewer APA...

Page 4: ...ble To record video the video cable may be connected to a suitable recording device The Video Viewer cannot be autoclaved and must be cleaned and disinfected between first use and each patient use refer to Section 4 The Handle incorporates and is powered by a single replaceable alkaline or lithium AA 1 5V battery and the Camera Module incorporates a high intensity light source and a miniature came...

Page 5: ...n a non sterile reusable state It will be necessary to clean it before use refer to Section 4 Check the battery of the laryngoscope Before using the device as a standard laryngoscope ensure that the Camera Module illuminates when the device locks in the ON position CAUTION The AA 1 5V battery MUST be changed regularly Inadequate battery level may cause the laryngoscope to fail suddenly without war...

Page 6: ...place until the limit of travel is reached Figure 3 3 To confirm correct attachment rotate the Camera Module into the ON position the Video Viewer screen should then turn on If using as a video laryngoscope the Video Viewer should be attached to the laryngoscope before using on a patient CAUTION If an image cannot be seen on the Video Viewer check that it is correctly attached to the Handle Avoid ...

Page 7: ... the epiglottis is elevated by lifting the Handle of the laryngoscope exposing the laryngeal inlet With direct visualisation of the laryngeal inlet a correctly sized ETT EndoTracheal Tube can be introduced into the trachea Ensure that the ETT is adequately lubricated prior to use An introducer stylet may be used to assist the introduction Correct placement of the ETT should be both visual and by t...

Page 8: ...t With direct visualisation of the laryngeal inlet a correctly sized ETT EndoTracheal Tube can be introduced into the trachea Ensure that the ETT is adequately lubricated prior to use An introducer stylet may be used to assist the introduction Correct placement of the ETT should be both visual and by the use of capnography Figure 3 7 Introducing the device with MIL Blade Figure 3 8 Connecting the ...

Page 9: ... fitted with the charger indefinitely if desired To begin charging attach the appropriate plug adapter to the charger Figure 4b and connect to the mains power supply Once connected align the convex point of the Video Viewer connector with the concave point of the Video Viewer Charger connector to confirm correct charging direction Figure 4c Attachment is made easy by the magnetic pins in the conne...

Page 10: ...the battery level is sufficient If the LED light indicator starts to flash during a procedure it means that there is limited battery The procedure may continue at the clinician s discretion The Camera Module is supplied with an alkaline AA 1 5V battery When used as a standard laryngoscope the Camera Module illuminates when the device locks in the ON position The battery should be replaced regularl...

Page 11: ...sed bag for transportation to the sterile processing unit for sterilisation CAUTION Only the Handle and Camera Modules may be sent for reprocessing Prepare or dilute a suitable disinfectant in accordance with the manufacturer s instructions Visit the U S Environmental Protection Agency EPA website for a list of registered disinfectants Spray or wipe the components with a suitable disinfectant i e ...

Page 12: ...battery If the illumination still does not work contact the local distributor for servicing refer to Section 5 5 Service and maintenance 6 Troubleshooting 6 1 As a video laryngoscope 6 2 As a laryngoscope The video laryngoscope should be inspected at regular intervals for Visible signs of external damage Scratches or defects on the Camera Module lens Build up of material on connection pin area of ...

Page 13: ...IN LIEU OF ALL OTHER WARRANTIES WRITTEN OR ORAL EXPRESS OR IMPLIED STATUTORY OR OTHERWISE NO IMPLIED STATUTORY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY AAM HEALTHCARE SHALL NOT BE LIABLE FOR ANY DAMAGES SUSTAINED BY BUYER ARISING FROM DELAY IN THE REPLACEMENT OR REPAIR OF PRODUCTS UNDER THE ABOVE WARRANTY Terms and conditions may change without prior notice Dif...

Page 14: ...g instructions for this product Residents outside the European Union must dispose or recycle this product in accordance with local laws or regulations that apply Device should be stored and used between 4ºC and 40ºC Keep dry Keep away from sunlight Date of manufacture Legal manufacturer EU and Manufactured for US Authorised representative in the European Community European Conformity mark Class 1 ...

Page 15: ...thcare All rights reserved AAM Healthcare APA and Venner are trademarks of the Venner Group of Companies This product is covered by patents and patent applications of the Venner Group of Companies in the US and other territories Visit www AAMHealthcare com intellectual property 2266 H January 2019 ...

Reviews: