background image

en

22

23

en

The following, is a list of possible problems and recommended solutions. 

If the problem persists, contact the local distributor.

Problem description

Possible cause

Action needed

Poor image quality  

(Image is not clear or dark)

Dust or fluid trapped in 

connector

Check cleanliness of the 

connectors

Video Viewer is not 

connected correctly with the 

laryngoscope

Check for firm connection 

between the devices

Illumination failure

Refer to Section 3

Contamination, fluid or 

lubricant on lens

Clean lens, refer to Section 4

No image  

(i) Screen turns off shortly 

after connection 

(ii) when devices are 

connected

(i) Video Viewer’s battery has 

no charge

Charge the Video Viewer, refer 

to Section 3.1

(ii) Video Viewer is not 

connected correctly with the 

laryngoscope

Check that the Video Viewer is 

inserted in the right direction 

and firmly connected with the 

Handle, refer to Section 2.4

Electronic circuitry fault

Return Video Viewer for 

servicing, refer to Section 5

Unable to turn on device 

even after long charging 

time

Video Viewer’s battery has 

exceeded usable life

Return Video Viewer for 

servicing, refer to Section 5

Electronic circuitry fault

Return Video Viewer for 

servicing, refer to Section 5

Video Viewer’s battery is not 

recharged after 2 hours

Video Viewer is not connected 

to its charger or the power 

supply

Check the connection to the 

charger and/or the power 

supply

Video Viewer is not inserted 

properly

Check for firm connection 

between the devices

If the Camera Module does not illuminate, check that the battery is inserted in the 

correct orientation or replace the battery with a new alkaline or lithium AA 1.5V battery.
If the illumination still does not work, contact the local distributor for servicing, 

refer to Section 5.

5. 

Service and maintenance

6. Troubleshooting

6.1 

As a video laryngoscope

6.2 

As a laryngoscope

The video laryngoscope should be inspected at regular intervals for:

  Visible signs of external damage
  Scratches or defects on the Camera Module lens
  Build-up of material on connection pin area of the Handle and Camera Module

Any other faults will require diagnosis by an authorised service centre, contact 

your local distributor or 

www.AAMHealthcare.com

 for technical support.

5.1 

General maintenance

The Video Viewer is recommended to be serviced every 2 years for a battery replacement. 
When returning the Video Viewer for servicing, pack the Video Viewer in its original 

product box or wrap the device carefully for protection. The device should be sent to the 

authorised service centre as advised by the local distributor.

5.1 

Preventive maintenance

In clinical situations where the risk of contamination and infection increases, use the  

APA IP Shield

, a disposable cover designed for the video laryngoscope, to further protect 

the device from these exposures.
Use the APA IP Shield

 VL if using the device as a video laryngoscope and  

APA IP Shield

 DL if using the device as a standard laryngoscope. Be sure to apply 

standard medical institution practices for using personal protective equipment during 

use. After use, the APA IP Shield

 should be disposed in accordance with standard 

departmental practices for handling contaminated items. Proceed to clean the  

APA

 Video Laryngoscope as per cleaning instructions, refer to Section 4.

CAUTION:

 

Always ensure that a used APA IP Shield

 is safely disposed prior to removing 

any personal protective equipment for the prevention of possible self-contamination.

4.5 

APA IP Shield

 VL and DL

Summary of Contents for APA Video Laryngoscope

Page 1: ...it ISTRUZIONI PER L USO ja 使用説明書 ko 사용설명서 nl GEBRUIKSAANWIJZING no BRUKSANVISNING pl INSTRUKCJA UŻYCIA pt INSTRUÇÕES DE UTILIZAÇÃO sv BRUKSANVISNING tr KULLANMA TALİMATI zh 使用说明 Venner Medical Singapore Pte Ltd 35 Joo Koon Circle Singapore 629110 www vennermedical com Advena Ltd Tower Business Centre 2nd Flr Tower Street Swatar BKR 4013 Malta www advenamedical com ...

Page 2: ...lade 15 2 6 5 APA Stylet 16 2 7 Removing the blade after use 16 3 Battery and charging information 17 3 1 Charging instructions for the Video Viewer 17 3 2 Video Viewer LED light indicators 18 3 3 Battery replacement in the Handle 19 4 Decontamination 20 4 1 Preparation 20 4 2 Cleaning 20 4 3 Disinfection 21 4 4 Sterilisation 21 4 4 1 Low temperature 21 4 4 2 Steam sterilisation 21 4 5 APA IP Shie...

Page 3: ...ryngoscopy in routine and difficult airway intubations It may be used to facilitate endotracheal intubation as part of general anaesthesia or CPR cardiopulmonary resuscitation during cardiac arrest It is suitable for use in pre hospital hospital and extended care facilities The video laryngoscope is reusable DEHP free and latex free It is fully detachable into three components APA Video Viewer APA...

Page 4: ...ble To record video the video cable may be connected to a suitable recording device The Video Viewer cannot be autoclaved and must be cleaned and disinfected between first use and each patient use refer to Section 4 The Handle incorporates and is powered by a single replaceable alkaline or lithium AA 1 5V battery and the Camera Module incorporates a high intensity light source and a miniature came...

Page 5: ...n a non sterile reusable state It will be necessary to clean it before use refer to Section 4 Check the battery of the laryngoscope Before using the device as a standard laryngoscope ensure that the Camera Module illuminates when the device locks in the ON position CAUTION The AA 1 5V battery MUST be changed regularly Inadequate battery level may cause the laryngoscope to fail suddenly without war...

Page 6: ...place until the limit of travel is reached Figure 3 3 To confirm correct attachment rotate the Camera Module into the ON position the Video Viewer screen should then turn on If using as a video laryngoscope the Video Viewer should be attached to the laryngoscope before using on a patient CAUTION If an image cannot be seen on the Video Viewer check that it is correctly attached to the Handle Avoid ...

Page 7: ... the epiglottis is elevated by lifting the Handle of the laryngoscope exposing the laryngeal inlet With direct visualisation of the laryngeal inlet a correctly sized ETT EndoTracheal Tube can be introduced into the trachea Ensure that the ETT is adequately lubricated prior to use An introducer stylet may be used to assist the introduction Correct placement of the ETT should be both visual and by t...

Page 8: ...t With direct visualisation of the laryngeal inlet a correctly sized ETT EndoTracheal Tube can be introduced into the trachea Ensure that the ETT is adequately lubricated prior to use An introducer stylet may be used to assist the introduction Correct placement of the ETT should be both visual and by the use of capnography Figure 3 7 Introducing the device with MIL Blade Figure 3 8 Connecting the ...

Page 9: ... fitted with the charger indefinitely if desired To begin charging attach the appropriate plug adapter to the charger Figure 4b and connect to the mains power supply Once connected align the convex point of the Video Viewer connector with the concave point of the Video Viewer Charger connector to confirm correct charging direction Figure 4c Attachment is made easy by the magnetic pins in the conne...

Page 10: ...the battery level is sufficient If the LED light indicator starts to flash during a procedure it means that there is limited battery The procedure may continue at the clinician s discretion The Camera Module is supplied with an alkaline AA 1 5V battery When used as a standard laryngoscope the Camera Module illuminates when the device locks in the ON position The battery should be replaced regularl...

Page 11: ...sed bag for transportation to the sterile processing unit for sterilisation CAUTION Only the Handle and Camera Modules may be sent for reprocessing Prepare or dilute a suitable disinfectant in accordance with the manufacturer s instructions Visit the U S Environmental Protection Agency EPA website for a list of registered disinfectants Spray or wipe the components with a suitable disinfectant i e ...

Page 12: ...battery If the illumination still does not work contact the local distributor for servicing refer to Section 5 5 Service and maintenance 6 Troubleshooting 6 1 As a video laryngoscope 6 2 As a laryngoscope The video laryngoscope should be inspected at regular intervals for Visible signs of external damage Scratches or defects on the Camera Module lens Build up of material on connection pin area of ...

Page 13: ...IN LIEU OF ALL OTHER WARRANTIES WRITTEN OR ORAL EXPRESS OR IMPLIED STATUTORY OR OTHERWISE NO IMPLIED STATUTORY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY AAM HEALTHCARE SHALL NOT BE LIABLE FOR ANY DAMAGES SUSTAINED BY BUYER ARISING FROM DELAY IN THE REPLACEMENT OR REPAIR OF PRODUCTS UNDER THE ABOVE WARRANTY Terms and conditions may change without prior notice Dif...

Page 14: ...g instructions for this product Residents outside the European Union must dispose or recycle this product in accordance with local laws or regulations that apply Device should be stored and used between 4ºC and 40ºC Keep dry Keep away from sunlight Date of manufacture Legal manufacturer EU and Manufactured for US Authorised representative in the European Community European Conformity mark Class 1 ...

Page 15: ...thcare All rights reserved AAM Healthcare APA and Venner are trademarks of the Venner Group of Companies This product is covered by patents and patent applications of the Venner Group of Companies in the US and other territories Visit www AAMHealthcare com intellectual property 2266 H January 2019 ...

Reviews: