33
EN
- After 5 seconds for Error messages Er1,
Temperature icon and Battery icon.
Dimensions:
107 mm (L) x 58 mm (W) x 13.5 mm (H).
Weight:
61 g (without batteries).
Meter operating
conditions:
- Temperature: 5 to 45 °C (41 to 113 °F) for glucose
testing; 10 to 40 °C (50 to 104 °F) for β-ketone
testing.
- Relative Humidity: 20 to 90% (no dew condensation)
for glucose testing; < 85% for β-ketone testing.
Meter storage
conditions:
- Temperature: -20 to 50°C (-4 to 122 °F).
- Relative Humidity: 20 to 90%.
Control Solutions
Storage Temperature:
4 to 30 °C (39.2 to 86 °F).
Operating
environment:
The meter complies with applicable electromagnetic
emission requirements (EMC).
Operating radio
frequency band(s):
13.56 MHz.
Maximum radio frequency
power transmitted in the
operating frequency band(s):
0.67 nW.
EU directive/
classification:
98/79/EC / Annex II, List B.
Device standard:
EN ISO 15197:2015.
Glucose measurement specifications
Unit of measurement:
mmol/L.
Test range:
1.1 - 33.3 mmol/L.
Haematocrit range:
10 - 70% (hæmatocrit compensated).
Test strip:
GLUCOFIX
®
TECH Sensor.
Sample size:
Minimum 0.5 μL.
Test time:
5 seconds.
Assay method:
Electrochemical, glucose oxidase (GOD,
Aspergillus
niger sourced) based method.
Mediator: Hexacyanoferrate(III) ion.
10. T
echnical specifications