7.6 NFC setup
The NFC function on your GLUCOFIX
®
TECH GK meter is preset to OFF and can be
enabled as follows:
1. Press the
▲
or
▼
buttons to select ON or OFF.
2. Press to confirm and exit.
8. METER CARE
8.1 Storing your meter
After use, tightly close the caps of the test strip vial and control
solution bottle
to maintain their quality.
Pack your meter, test strips, control solutions and manuals into your carry case
and store in a dry place. The correct storage temperature is -20 to 50 °C (-4 to
122 °F) for the meter and 4 to 30 °C (39.2 to 86 °F) for the test strips and control
solutions. Do not freeze. Avoid heat, humidity and direct sunlight.
CAUTION
To obtain accurate test results:
●
Do not use test strips or control solutions if their bottles or foil pouch are
broken or have been left open.
●
Do not use the test strips or control solutions after their expiry dates.
8.2 Cleaning your meter
Your meter does not need special cleaning. If your meter gets dirty, wipe it with
a soft cloth moistened with mild detergent. To disinfect your meter after
cleaning, wipe with a soft cloth moistened with either 75% ethanol alcohol or
diluted household bleach (10% sodium hypochlorite solution).
CAUTION
●
DO NOT get water inside the meter. Never immerse the meter or hold it
under running water.
●
DO NOT use glass cleaners or household cleaners on the meter.
●
DO NOT try to clean the test strip port.