33
32
Instrument Settings
Preprogrammed operators
•
‘autologin’: See “11.4.2 System security settings”
•
‘self add’: See “11.4.2 System security settings”
•
‘supervisor’: Supervisor-level operators can modify system security settings.
The operator name is ‘supervisor’ (all lower-case, without the inverted com-
mas), and the default password is ‘1234’. Supervisor-level operators can never
be displayed on the
Login
screen.
•
‘service’: Service-level operators can access the
Service menu
screen.
•
‘Full database and config clear.’: If this string is entered (as is, without the
inverted commas, but with a capitalized first word and a period (full stop) at
the end) as an Operator name on the
Login
screen, the system will perform a
full database clear.
ǽ
Full clear is a final, irrevocable command. Use it only when necessary. It is
recommended to perform a “Log export 255” before a Full clear.
10.4.3 Security settings overview
1
Open
system
2
Anonymous
usage
3
Self-add
4
Self-
add with
password
5
Secure
auto login
On
On
Off
Off
Off
auto login
rights
admin
user
N/A
N/A
N/A
self add
Off
On
On
On
Off
self add rights N/A
user
user
user
N/A
password not
required
On
On
On
Off
Off
perform test
anyone
(anonymous)
anyone
(anonymous)
anyone
anyone
registered
users
modify
settings
anyone
admins
admins
admins
admins
modify
security
supervisor
(def
password)
supervisor
(def
password)
supervisor
(def
password)
supervisors
supervisors
add user
N/A
anyone
anyone
anyone
admins
login
autologin
autologin
self-
registered
users w/o
passw
self-
registered
users with
passw
admin-
registered
users with
passw
user
management
N/A
admins
admins
admins
admins
identification
not forced
not forced
forced
forced
forced
password
usage
not forced
not forced
not forced
yes
yes
real audit trail
no
no
no
yes
yes