3M™ Versaflo™ Powered and Supplied Air Respiratory Systems
Additional Specifications
Respirator System
Operating Temperature
•
23º F to 129
°
F (-5 to 54
°
C).
•
Motor blower battery alarm will activate when battery pack
p
g
p
Operating Altitude
Range
y
y p
internal temperature exceeds 130º F (55º C).
•
Motor blower will shut down if the battery pack temperature
reaches 140
°
F (60
°
C).
Storage Temperature
(RH <90%)
TR-300 motor blower,
HE filter Battery pack
•
-22º F to 122ºF (-30 to 50
°
C).
•
Note: 40ºF to 95º F (4º C to 35º C) suggested storage.
temperature range if product will be stored for an extended
period of time before first use
HE filter, Battery pack
period of time before first use.
•
Optimal storage temperature for the battery pack is 59º F
(15º C).
Shelf Life
•
Motor/blower: 5 years
(
If motor blower will not be used on a regular
y
(
g
basis, the motor must be run for 5 minutes annually to ensure proper
lubrication and operation).
•
Battery pack: 9 months.
•
High Efficiency (HE) filter: 5 years.
8
© 3M 2011. All Rights Reserved. Rev 1
g
y ( )
y
•
Prior to use, when stored in original 3M packaging.