Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document.
www.rangerfluidwarming.com reorder #202524A
e
xplanatIon
of
S
Ignal
w
ord
c
onSequenceS
WARNING:
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION:
Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE:
Indicates a situation which, if not avoided, could result in property damage only.
w
arnIngS
and
c
autIonS
WARNINGS:
1. To reduce the risks associated with hazardous voltage and fire and thermal energy hazards:
• Do not substitute other devices (i.e. Model 245) for the Ranger irrigation fluid warming unit
or Ranger irrigation warming sets.
2. To reduce the risks associated with hazardous voltage and fire:
• Do not modify this device.
• Connect power cord to receptacles marked “Hospital Only,” “Hospital Grade,” or a reliably
grounded outlet.
• Use only the power cord specified for the product and certified for the country of use.
• Do not allow the power cord to get wet.
• Do not use the warming unit when it appears the Ranger warming unit, power cord, or
warming set is damaged. Use only 3M specified replacement parts.
• Keep power cord visible and accessible at all times. The plug on the power cord serves as
the disconnect device. The wall socket outlet shall be as close as practical and shall be easily
accessible.
CAUTION:
1.
To reduce the risks associated with cross-contamination:
• The cleaning tool provides only superficial cleaning, it does not disinfect or sterilize the
interior of the unit.
2.
To reduce the risks associated with impact and facility medical device damage:
• Clamp the Ranger warming unit to an I.V. pole with a minimum 14" (35.6 cm) radius
wheelbase and at a height no higher than 44" (112 cm).
3.
To reduce the risks associated with environmental contamination:
• Follow applicable regulations when disposing of this device or any of its electronic
components.
4.
Do not use the Ranger blood/fluid warming system 245 Model for direct cardiac application.
When using the Model 245 with a central venous catheter (CVC), ensure that the catheter tip
does not have direct contact with the heart and ensure that all electrical devices connected to
or near the patient have an appropriate leakage current rating for the application. If a CVC tip
is found to be in direct contact with a patient’s heart, the Model 245 should be immediately
disconnected from the CVC until the CVC is safely repositioned. Failure to follow these
precautions may cause cardiac disturbance and/or patient injury.