18 Specifications
Performance characteristics
Operating pressure
300 mmHg setpoint
Note:
• Pressure system is In Range when the pressure infusor bladders are inflated to between 230 mmHg (low) and 330 mmHg
(high). If pressure falls below 230 mmHg for more than approximately 30 seconds the Low amber indicator will
illuminate and an audible alarm will sound. The High amber indicator and an audible alarm notify the user when the
pressure infusor bladder is above 330 mmHg.
• The outlet pressure of the fluid may vary with the surface area and volume of the solution bag.