Introduction
Product description
The Ranger pressure infusor can be used with the Ranger blood/fluid warming unit, standard and high flow disposable sets, the Ranger
irrigation warming unit, and irrigation warming sets
1
. The infusor can accept solution bags ranging from 250 mL to 1000 mL. The
pressure infusor delivers a maximum fluid pressure of 300 mmHg.
Note: The solution pressure of the Ranger pressure infusor is dependent on surface area and volume of the solution bag. To verify the
pressure, refer to the Maintenance Guide (part # 210393).
The Ranger pressure infusor has no user-adjustable controls. The user slides an I.V. solution bag behind the metal fingers and against
the inflation bladder located inside the pressure infusor.
When the infusor is attached to an external power source and the main power switch is turned ON, pushing the pressure infusor
power button ON inflates the inflation bladder and maintains pressure on blood and solution bags.
Turning the pressure infusor power button OFF deflates the bladder. Turn the main power switch OFF when the pressure infusor is not
in use.
1 Use a maximum of 1 liter solution bags when using the Ranger pressure infusor with the Ranger irrigation fluid warmer.
Power entry module
Main power switch