7.  Monitor Patient Heart Rate

The Model 3100 detects and displays an acoustic-based heart rate when presented with consistent heart sounds (heart rate 
variation < 10%). It takes fi ve seconds to compute the initial heart rate and updates are provided every two seconds. Prior 
to the initial reading, the display shows two dashes (--). For heart rates outside a range of 30-199 bpm, the display will also 
show two dashes (--). 

The acoustic-based heart rate display functions best when the Model 3100 is placed near the apex of the patient’s heart and 
can be monitored while using any fi lter mode and/or volume level. If the heart rate changes from consistent to inconsistent or 
if there is excessive ambient noise, patient movement or lung sounds during auscultation, the heart rate display number will 
display two dashes (--). 

 

 

8.  Monitor Battery Life

Battery life is indicated by an icon in the LCD display. 

 

The Model 3100 comes with a AA Alkaline battery. This battery will last for approximately 60 hours of continuous use. In a 
typical clinical setting, this represents about three months. 

As the alkaline battery life depletes, the icon will change as follows:

 

The battery icon begins to blink when only a few hours of battery capacity remains. 

 

CAUTION:

 When the battery is completely depleted the scope becomes inoperable. No sound will be transmitted without 

a functioning battery.

IMPORTANT! 

NiMH (rechargeable) and Lithium batteries may also be used in the Model 3100. However, battery type must be 

specifi ed to ensure a reliable battery life indication (see 

Custom Confi guration

 section below). 

9. Custom 

Confi guration

The Model 3100 stethoscope has several operational settings that can be changed using its confi guration menu system. 
Within this menu system, the (+) and (-) buttons are used for scrolling and the (M) button is used for selecting options. The 
Filter button acts an “escape” key, used for returning to the previous menu or to the operating mode of the stethoscope.

To enter the Setup Menu: 

Depress and release the (M) button. This will display a menu of setup options. (NOTE: Only 3 

options will visible at a time. All options can be viewed in sequence upon scrolling with either the (+) or (-) buttons.)

 

To select the power setting: 

Select the “POWER” option in the Setup menu. Use the (-) and (+) buttons to scroll through the 

list of power settings. Press and release the (M) button to select a setting.

 

Summary of Contents for Littmann 3100

Page 1: ......

Page 2: ...this latest electronic stethoscope from the Littmann brand Hear it With Confidence SAFETY INFORMATION Please read understand and follow all safety information contained in these instructions prior to...

Page 3: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Page 4: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Page 5: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Page 6: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Page 7: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Page 8: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Page 9: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Page 10: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Page 11: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Page 12: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Page 13: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Page 14: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Page 15: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Page 16: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Page 17: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Page 18: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Page 19: ......

Reviews: