Declaration – Electromagnetic Emissions

The 3M Littmann® Electronic Stethoscope, Model 3100, is intended for use in the electromagnetic environment specifi ed 
below. The customer or the user of the Model 3100 should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment –guidance

RF emissions CISPR 11

Group 1

Model 3100 uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class B

Model 3100 is suitable for use in all establishments, including 
domestic establishments and those directly connected to 
the public low-voltage power supply network that supplies 
buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2

Not applicable

Voltage fl uctuations/fl icker emissions 
IEC 61000-3-3

Not applicable

Declaration – electromagnetic immunity

The 3M Littmann® Electronic Stethoscope, Model 3100, is intended for use in the electromagnetic environment specifi ed 
below. The customer or the user of Model 3100 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment –guidance

Electrostatic discharge 
(ESD) IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or 
ceramic tile. If fl oors are covered with 
synthetic material, the relative humidity 
should be at least 30%.

Electrical fast transient/
burst IEC 61000-4-4

± 2 kV for supply lines
± 1 kV for input/output lines

Not applicable

Surge IEC 61000-4-5

± 1 kV differential mode
± 2 kV common mode

Not applicable

Power frequency 
(50/60 Hz) magnetic 
fi eld IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fi elds should 
be at levels characteristic of a typical 
location in a typical commercial magnetic 
fi eld or hospital environment.

Voltage dips, short 
interruptions and 
voltage variations on 
power supply lines IEC 
61000-4-11

< 5 % UT (>95% dip in UT) for 
0.5 cycle
40 % UT (60% dip in UT) for 
5 cycle
70 % UT (30% dip in UT) for 
25 cycle
< 5% UT (>95% dip in UT) 
for 5 sec

Not applicable 

Summary of Contents for Littmann 3100

Page 1: ......

Page 2: ...this latest electronic stethoscope from the Littmann brand Hear it With Confidence SAFETY INFORMATION Please read understand and follow all safety information contained in these instructions prior to...

Page 3: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Page 4: ...ope Read understand and follow all the safety information on the battery package Intended Use The 3M Littmann Electronic Stethoscope Model 3100 is intended for medical diagnostic purposes only It may...

Page 5: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Page 6: ...o during standby mode When stethoscope diaphragm or any button is pressed LCD display settings will reappear and stethoscope will be ready to use After two hours of non use the stethoscope will leave...

Page 7: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Page 8: ...hs As the alkaline battery life depletes the icon will change as follows The battery icon begins to blink when only a few hours of battery capacity remains CAUTION When the battery is completely deple...

Page 9: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Page 10: ...OW setting uses less battery power To increase battery life ENABLE the Auto Off feature and select the LOW power setting The power settings provide different degrees of power consumption In general th...

Page 11: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Page 12: ...ering off Backlight remains lit for 5 seconds after pressing power button MEDIUM Remains on full power for 5 minutes before entering standby Remains in standby for 2 hours before powering off Backligh...

Page 13: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Page 14: ...Cleaning Other Parts of the Stethoscope Eartips eartubes plastic tubing and chestpiece can be wiped clean with alcohol Eartips may be removed for a more thorough cleaning NOTICE Do not immerse the ste...

Page 15: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Page 16: ...e 3M Littmann Electronic Stethoscope Model 3100 is intended for use in the electromagnetic environment specified below The customer or the user of Model 3100 should assure that it is used in such an e...

Page 17: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Page 18: ...accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Model 31...

Page 19: ......

Reviews: