Starting Position
Rotated Position
The Rotation Indicator gives you real-time, on-screen feedback about the orientation of the
wand. It is a visual cue to help you determine how the wand is rotated compared to its
starting, occlusal view orientation. In the example above, the starting position is facing the
occlusal surface; the rotated position is nearly 90 degrees on the buccal side.
Note: The Rotation Indicator does not display when the wand is held vertically.
For more information, refer to your Lava C.O.S. training materials.
Maximum Scanning Times
There are limits to the scanning times for a case. The Lava C.O.S. will stop a scan if:
•
the current continuous scan exceeds 7 minutes
•
the current continuous bite scan exceeds 2 minutes
•
the total of the current case’s maxillary scans exceeds 7 minutes
•
the total of the current case’s mandibular scans exceeds 7 minutes
•
the total of the current case’s bite scans exceeds 6 minutes
•
the total of the current case’s additional scans exceeds 4 minutes
•
the total of all scans (not including bite scans) exceeds 15 minutes
To avoid reaching these maximums, it’s a good practice to delete any unnecessary or
superfluous scans for a case.