15.2 Device Handling TIPS
Recommendations
Operate and store the device in dry conditions. Do not operate or store the device in damp surroundings for a long period of time.
Operate the device within the defined pressure range. Do not exceed a test pressure of 60 psi (4,137 mbar). The device has a maximum
operating pressure of 65 psi (4,500 mbar).
Operate the device within the defined temperature range. Operating outside the defined temperature range leads to erroneous measure-
ment results or error messages.
Connect to the filter system only with standard compressed air couplings, matching the nipple, depending on the design of the device.
Changes to the air pressure nipple of the device can lead to leaks, damages to the battery pack or data loss.
Charge the device at least every two months. Storing for more than two months without a charge can lead to damages to the battery pack
or data loss.
Use original power adapter. Use of a power supply different from the original power supply can cause damages to the device.
Caution: If the device is stored above 30 °C for a long period of time, the life of the battery pack will be reduced, even if the device is
switched off. In order to avoid data loss, charge the device more often.
15.3 Maintenance Plan
Part of device/system
Maintenance instruction
Interval
Action
General
Visual inspection
At least prior to every use
If dirty, clean with a damp cloth
Battery
Charge battery
At least every 30 days
—
Pressure sensor
Calibrate pressure sensor
At least every 12 months
—
15.4 Spare Parts Ordering List
Parts
3M #
3M™ 101 Series Integrity Test Device Int’l Charger/Power supply cord
adapters
70020334440
3M™ 101 Series Integrity Test Device Battery pack
70020334457
USB/USB-mini cord
70020334465
ABS black case
70020334473
3M™ 101-Series Integrity Test Device Op-Manual
34-8719-8229-3
Printer Charger/Power supply cord
70020334507
Printer Battery pack
70020334515
Thermo print paper, roll
70020334523
PrinterOp-Manual
70020334531
-24-