19196-001 Rev. D
VIDAR Advantage Series User’s Guide
8
DosimetryPRO
®
Advantage specifications
Nominal
resolution
Pixels
(14"x17" film)
Spot size
(µm)
DPI
Line pairs
per mm
1k x 1.25k
997 x 1211
356
71.25
1.4
2k x 2.5k*
1995 x 2422
178
142.5
2.8
4k x 5k
3990 x 4845
89
285
5.6
*ACR Standard for Teleradiology [Revision 35 (1998)] recommends 2.5 lines pairs/mm minimum
Scan modes
32-bit mapped to 12-bit and 8-bit grayscale outputs
Film sizes
Manual
Feed:
Width:
6" to 14" (15.24cm to 35.56cm)
Length:
7" to 51"† (17.78cm to 129.54cm)
Thickness: 0.005" to 0.009" (0.127mm to 0.228mm)
†Films longer than 17” require user support during feeding, and a scanning application that handles long
films
Auto film feeder
Standard 25-film capacity (mixed sizes, no presorting)
“Light box” loading: head-up, normal reading, left justified
Width:
7" to 14" (17.78cm x 35.56cm)
Length
7" to 17" (17.78cm x 43.18cm)
Thickness: 0.005" to 0.009" (0.127mm to 0.228mm)
Note:
ISP 14" x 17" is cut true to size and is therefore slightly
larger than standard 14" x 17" film (which is actually measured
in metric and therefore slightly smaller). This results in a very
tight fit between the digitizer’s film guides. In most cases, the
films can be manually loaded and may need to be assisted
through the device. In the worst case, the film may need to be
slightly trimmed before feeding. Films will not be properly fed
using the automated film feeder.