5.
Carefully retrieve the new lamp module from its transporting container. Hold the lamp
module only by its ceramic part.
6.
Insert the new lamp module. Make sure that the groove (
1
) in the contact surface of the
lamp captures the corresponding pin of the lamp mount (see
Figure 80
).
Figure 80 – Inserting the lamp module
Caution!
Incorrect handling may considerably reduce the lamp
module’s useful
life. Hold it by the ceramic part in all cases! Never touch the lamp’s
glass surface or the inside surface of the reflector. Skin oil is harmful to
these parts.
7.
Tighten pressure bolt again with the lever (
Figure 79 – 1
).
8.
Re-establish plug-in connection.
9.
Screw the lateral cover-plate down to the housing.
10.
Connect power plug.
11.
Set hour meter (time counter) to zero.
5.3.4 Removing a slide
If a slide got jammed or has suffered breakage during the process of digitization, stop the process
and, if possible, move the slide holder to a position where the slide can be easily removed by clicking
the
Emergency Manual Slide Removal Position
).
May 29, 2014 - Rev. 1
3DHISTECH Ltd.
117
(