4 Software Description
Figure 70 – Scan To Server window in Automatic Mode
•
Give the number of the starting (
First slide
) and the final slide (
Last slide
) to be
scanned in the appropriate field. For example, if you want to scan slides to server
from 1-12, type: First slide:
1
; Last slide:
12
.
•
There are three options at the bottom of the window:
1. Server side barcode parsing
option can be selected if CaseCenter is connected
with the supported HIS/LIS. This way the barcode read by the software is being
identified, and the predefined location of that specific case in CaseCenter is
automatically selected as scanning destination based on the CaseID (containing
slide and patient data stored on HIS/LIS). Therefore,
Slide name
,
Folder
,
Case
,
and
Block
field data cannot be modified, only a profile can be selected.
2.
The
Generate slide name if no name was defined
option can be activated when
you do not want to type in a unique name for each slide.
3.
Upon activating the
Read barcode if no name is defined on the list
option, the
slide barcode will be read and set in the Barcode field.
•
Use masks for automatic fills. Activate the
Use mask
option and the mask fields
turn fillable. To save masks, click the
Save Mask
button. To modify masks click
Set
Mask
. For example, by creating a string for the slide name fields based on the
characters of the barcode, a unique name will be generated. The barcode
'0123456789', and a mask '1-3' result that the first four characters of the barcode
will be displayed and set as slide name (012).
Case
and
Block
fields can be set this
way as well, if necessary (see
Figure 71
).
NOTE:
When filling slide data fields, press
TAB
to move to the next cell.