background image

5.2.5 DELETING OF O

2

 MEASURED  VALUES  

  

FROM MEMORY

 

The  measured  values  can  be  deleted  from  memory  any  time.  The  memory  capacity  will  be  completely  recovered 
(450 values) and newly measured values will be saved  in sequence   from beginning of the memory.  
 
1) Press the key ON to switch on the device. The text „1-CUBE“ with the program version appears on the display for 
a moment.   Then the following text appears: „

ENT.meas.

“ 

                                                                            „

+ view.

“ 

2) Press the key mínus (-).Then the following text appears  „

memory

“ 

                                                                                     

deleting“

   

And all the saved values are  overwritten with number   65535  (free capacity of the memory).   
 

 
5.2.6 REPLACEMENT OF BATTERIES 
 

1)

 

unscrew as shown below 

 

2)

 

Pull the element out and replace batteries as shown below 

 

Summary of Contents for OXI

Page 1: ...GEN MEASURING DEVICE USER S GUIDE Contents 1 Range of supply and device application 2 Installation putting the device into operation 3 Safety recommendations 4 Technical data 5 Operating instructions 6 Service changing the O2 cap refuse disposal 7 Installation of communication program into PC and measured data transfer ...

Page 2: ... of inlet hose with the connection 4 on oximeter see pic 2 to the sampling point In case it is not possible to connect the inlet hose to the sampling point please use the corresponding reduction attach the corresponding hose to the reduction to take safely the sample The inlet hose 3 must be secured at the sampling point not to fall out during taking of sample Connect the reduction 1 with the help...

Page 3: ...rages and gases with manfacturer Check device before each measurement Do not use visibly damaged device Notice Remove sludge from the sampling point before connecting the device Remove sludge of the sedimented yeast to fill inside the chamber only beer without sedimented yeast Before the measurement connect the device by fixing of inlet hose with the connection 4 on oximeter see pic 2 to the sampl...

Page 4: ... water cock Disconnect the device from the water cock and let the water go out from the measuring chamber Clean the external surface with soft flannel moistened with water After this handling the device is ready for next measurement 5 2 CONTROL OF THE ELECTRONICS The device i e switch on shutdown measurement saving of the measured sample in the memory is controlled with the help of the membrane ke...

Page 5: ... the running period and the accumulator life time will be shortened too Once the charging of the device is over unplug the adaptor and the opposite connector of the adaptor from the connector on the device At the end screw fully the connector cover on the device 5 2 3 O2 CONTENT MEASUREMENT Follow the paragraphs 5 1 1 then further 1 Press the key ON to switch on the device The text 1 CUBE with the...

Page 6: ...mory 1 Press the key ON to switch on the device The text 1 CUBE with the program version appears on the display for a moment Then the following text appears ENT meas view 2 Press the key mínus Then the following text appears memory deleting And all the saved values are overwritten with number 65535 free capacity of the memory 5 2 6 REPLACEMENT OF BATTERIES 1 unscrew as shown below 2 Pull the eleme...

Page 7: ...com At the end of the apparatus life we recommend to follow the Waste Act The electronic parts has to be destroyed separately by a company dealing with electronic waste disposal Take the metallic parts made of iron brass plastic elements and packing material to the separated collection as a secondary raw material You can also order the apparatus liquidation at the above mentioned company ...

Page 8: ...6 1 CHANGING THE O2 SENSOR CAP Loosen the connector 7 and unscrew O2cap 5 from the measuring chamber 3 see pic 2 ...

Page 9: ...6 8 9 Legend 1 Outlet hose reduction 2 Control valve 3 Measuring chamber 4 Intlet hose reduction 5 O2 Electrode 6 Keyboard with display see pic 1 7 O2 Electrode Connector 8 Oximeter body with microcomputer 9 Connector input see pic 4 ...

Page 10: ...Pic 3 Set Oxi hopped wort measurement 1 ICAS 2 4 Legend 1 Outlet hose reduction 4 Intlet hose reduction ...

Page 11: ... 1 CUBE with the program version appears on the display for a moment Then the following text appears memory view 4 Press the key plus There is text memory on the first line and the measured O2 values appear one by one from the oldest to the newest ones on the second line At the same time these values are displayed in SerialWatcher program Windows in column one by one from the oldest value to the n...

Reviews: