Pic. 4  Connector input for connection of the charging adapter and the measured data 
transfer into PC 

 

 

            

 

 

6.2  O

2   

MEASURED  VALUES  TRANSFER    FROM  THE  MEMORY  INTO  PC  (ONLY 

OXI COMFORT TYPE) 
 

Possibility of data  transfer through comunnicating cable into PC. Before the first data transfer from the device into 
PC,  install program SerialWatcher  from our supplied CD.  
 
1)  Connect  the  shut-off  device    OXI  comfort    with  the  help  of  cable    FTDI  USB-RS485-WE-1800-BT    (supplied 
with the device    OXI comfort) with  USB port of your PC. 
2) Run the program  SerialWatcher in your PC  and set the corresponding COM port (f.e..COM 5), speed of transfer  
     9600  and  Display ASCII. 
3) Press the key ON to switch on the device OXI comfort .  The text „1-CUBE“ with the program version appears on 
the display for a moment.   Then the following text appears: „

memory

“ 

                                                                                        

view“

   

4) Press the key plus (+).There is text „

memory

“ on  the  first  line  and  the  measured O

2

 values appear one by one 

from the oldest  to the newest ones  on the second line. 
At the same time  these values are displayed  in SerialWatcher program  Windows in  column  one by one  from the 
oldest  value to the newest ones. These values can be saved as file txt  under any chosen name. 
     Notice. The measured values are displayed in ppb in SerialWatcher program.   

 

6.3 Calibration

 

The device is calibrated at 2 points  (0ppb and air). Calibration is run from PC  with program 

Hamilton Device 

Manager 1.0.0.. 

The program can be uploaded for free at www.hamiltoncompany.com.  It is necessary  to have 

connector USB-RS485 ModBus  Converter.  

 
User’s Guide - Movie Links: 

http://www.youtube.com/watch?v=1tLYJnsVIDw

 

http://youtu.be/I6Vy4xH4ZDQ

 

 

Summary of Contents for OXI

Page 1: ...GEN MEASURING DEVICE USER S GUIDE Contents 1 Range of supply and device application 2 Installation putting the device into operation 3 Safety recommendations 4 Technical data 5 Operating instructions 6 Service changing the O2 cap refuse disposal 7 Installation of communication program into PC and measured data transfer ...

Page 2: ... of inlet hose with the connection 4 on oximeter see pic 2 to the sampling point In case it is not possible to connect the inlet hose to the sampling point please use the corresponding reduction attach the corresponding hose to the reduction to take safely the sample The inlet hose 3 must be secured at the sampling point not to fall out during taking of sample Connect the reduction 1 with the help...

Page 3: ...rages and gases with manfacturer Check device before each measurement Do not use visibly damaged device Notice Remove sludge from the sampling point before connecting the device Remove sludge of the sedimented yeast to fill inside the chamber only beer without sedimented yeast Before the measurement connect the device by fixing of inlet hose with the connection 4 on oximeter see pic 2 to the sampl...

Page 4: ... water cock Disconnect the device from the water cock and let the water go out from the measuring chamber Clean the external surface with soft flannel moistened with water After this handling the device is ready for next measurement 5 2 CONTROL OF THE ELECTRONICS The device i e switch on shutdown measurement saving of the measured sample in the memory is controlled with the help of the membrane ke...

Page 5: ... the running period and the accumulator life time will be shortened too Once the charging of the device is over unplug the adaptor and the opposite connector of the adaptor from the connector on the device At the end screw fully the connector cover on the device 5 2 3 O2 CONTENT MEASUREMENT Follow the paragraphs 5 1 1 then further 1 Press the key ON to switch on the device The text 1 CUBE with the...

Page 6: ...mory 1 Press the key ON to switch on the device The text 1 CUBE with the program version appears on the display for a moment Then the following text appears ENT meas view 2 Press the key mínus Then the following text appears memory deleting And all the saved values are overwritten with number 65535 free capacity of the memory 5 2 6 REPLACEMENT OF BATTERIES 1 unscrew as shown below 2 Pull the eleme...

Page 7: ...com At the end of the apparatus life we recommend to follow the Waste Act The electronic parts has to be destroyed separately by a company dealing with electronic waste disposal Take the metallic parts made of iron brass plastic elements and packing material to the separated collection as a secondary raw material You can also order the apparatus liquidation at the above mentioned company ...

Page 8: ...6 1 CHANGING THE O2 SENSOR CAP Loosen the connector 7 and unscrew O2cap 5 from the measuring chamber 3 see pic 2 ...

Page 9: ...6 8 9 Legend 1 Outlet hose reduction 2 Control valve 3 Measuring chamber 4 Intlet hose reduction 5 O2 Electrode 6 Keyboard with display see pic 1 7 O2 Electrode Connector 8 Oximeter body with microcomputer 9 Connector input see pic 4 ...

Page 10: ...Pic 3 Set Oxi hopped wort measurement 1 ICAS 2 4 Legend 1 Outlet hose reduction 4 Intlet hose reduction ...

Page 11: ... 1 CUBE with the program version appears on the display for a moment Then the following text appears memory view 4 Press the key plus There is text memory on the first line and the measured O2 values appear one by one from the oldest to the newest ones on the second line At the same time these values are displayed in SerialWatcher program Windows in column one by one from the oldest value to the n...

Reviews: