5.2.2 CHARGING OF THE DEVICE 

 
Fixed type of the device  (without  battery)  has to be  permanently connected to the network trough the adapter 24 
V DC/200 mA, which is connected with  oximeter.  
The information below is valid only  for the portable type of oximeter supplied with the rechargeable batteries. 

 

 

If the battery voltage is under 16,5 V  low  then after the switch-on the device the  text „Recharge battery“ appears 
on  the  display  for  a  second.  It  is  possible  to  work  with  the  device  for  another  30  minutes.  If  the  device  is  not 
recharged during this time then the  text „Low voltage“ appears on the display and the device is shut down.  
If you want to monitor how the battery is charged while working with the device (when it doesn’t measure O2) – then  
press  the  rotation key -  the  text „battery)“ appears on the first line of display  and current voltage on the second 
line of display. Unscrew the connector cover  and put the opposite  connector of the adaptor in to  charge the device. 

Warning:

 wach the right orientation (bump  in the adapter connector into the adapter slot. Boss and slot are 

marked with coloured dot)

.

 Adaptor can be plugged in el. socket 230V, 50Hz  now. During charging of the device 

the led BAT is shining. 

Notice: In case the connectors will not be put in the right way they will be damaged. This type of damage is 
not accepted by manufacturer under warranty. 

 

The device can be charged only by the supplied adaptor (output voltage of adapter is 24 V DC/200 mA). Average  
charging takes about 8 hrs.  I n case of need the period can be shortened but the running period and the accumulator 
life-time  will  be  shortened  too.  Once  the  charging  of    the  device  is  over    unplug    the  adaptor  and    the  opposite  
connector of the adaptor from the connector on the device.  At the end screw fully the connector cover on the device.  
 

5.2.3 O

2

 CONTENT MEASUREMENT   

 
Follow the paragraphs   5.1.1. then further: 
 
1) Press the key ON to switch on the device. The text „1-CUBE“ with the program version appears on the display for 
a moment.   Then the following text appears: „

ENT.meas.

“ 

 

                                                      „

+ view.

“  

 

2) Press the key ENTER. The already measured value of O2 appears on the display  - on the first line in ppb,   
on the second line in ppm for  example: 

„ppb    1253“

 

                                                            „ppm 1,253“ 

The displayed values are saved in ppb units.  
 
Press the key ENTER to finish the measurement, the text: „

ENT.meas.

“ 

                                                                                           „

+ view.

“ 

 

 

And the device is ready for next measurement  
 

5.2.4 SCANNING MEMORY OF THE MEASURED VALUES O

2  

 

The measured values of O2 are  saved in the device memory  even after its shutdown. The capacity of memory is 450 
of measured values.  
1) Press the key ON to switch on the device. The text „1-CUBE“ with the program version appears on the display for 
a moment.   Then the following text appears„

ENT.meas.

“ 

 

                                                            „

+ view.

“ 

 

 

2) Press the key plus (+).There is text „

memory

“ on the first line and the measured values of O

2

   in    ppb    appear 

one by one from the oldest value to the newest ones  on the second line. After the last saved value the numbers start 
to  appear      65535    (free  capacity  of  the  memory).      Press  the  key  ENTER  to    stop  scanning  memory  Then  the 
following text appears„

ENT.meas.

“ 

                                         „

+ view.

“ 

 

Notice. Once there is  the full storage, the memory will be automatically deleted.   
Then the following text appears: 
 „

memory

“ 

 

deleting“

   

 And all the saved values are  overwritten with number   65535  (free capacity of the memory).   
 
  
 
 
 
 
 

Summary of Contents for OXI

Page 1: ...GEN MEASURING DEVICE USER S GUIDE Contents 1 Range of supply and device application 2 Installation putting the device into operation 3 Safety recommendations 4 Technical data 5 Operating instructions 6 Service changing the O2 cap refuse disposal 7 Installation of communication program into PC and measured data transfer ...

Page 2: ... of inlet hose with the connection 4 on oximeter see pic 2 to the sampling point In case it is not possible to connect the inlet hose to the sampling point please use the corresponding reduction attach the corresponding hose to the reduction to take safely the sample The inlet hose 3 must be secured at the sampling point not to fall out during taking of sample Connect the reduction 1 with the help...

Page 3: ...rages and gases with manfacturer Check device before each measurement Do not use visibly damaged device Notice Remove sludge from the sampling point before connecting the device Remove sludge of the sedimented yeast to fill inside the chamber only beer without sedimented yeast Before the measurement connect the device by fixing of inlet hose with the connection 4 on oximeter see pic 2 to the sampl...

Page 4: ... water cock Disconnect the device from the water cock and let the water go out from the measuring chamber Clean the external surface with soft flannel moistened with water After this handling the device is ready for next measurement 5 2 CONTROL OF THE ELECTRONICS The device i e switch on shutdown measurement saving of the measured sample in the memory is controlled with the help of the membrane ke...

Page 5: ... the running period and the accumulator life time will be shortened too Once the charging of the device is over unplug the adaptor and the opposite connector of the adaptor from the connector on the device At the end screw fully the connector cover on the device 5 2 3 O2 CONTENT MEASUREMENT Follow the paragraphs 5 1 1 then further 1 Press the key ON to switch on the device The text 1 CUBE with the...

Page 6: ...mory 1 Press the key ON to switch on the device The text 1 CUBE with the program version appears on the display for a moment Then the following text appears ENT meas view 2 Press the key mínus Then the following text appears memory deleting And all the saved values are overwritten with number 65535 free capacity of the memory 5 2 6 REPLACEMENT OF BATTERIES 1 unscrew as shown below 2 Pull the eleme...

Page 7: ...com At the end of the apparatus life we recommend to follow the Waste Act The electronic parts has to be destroyed separately by a company dealing with electronic waste disposal Take the metallic parts made of iron brass plastic elements and packing material to the separated collection as a secondary raw material You can also order the apparatus liquidation at the above mentioned company ...

Page 8: ...6 1 CHANGING THE O2 SENSOR CAP Loosen the connector 7 and unscrew O2cap 5 from the measuring chamber 3 see pic 2 ...

Page 9: ...6 8 9 Legend 1 Outlet hose reduction 2 Control valve 3 Measuring chamber 4 Intlet hose reduction 5 O2 Electrode 6 Keyboard with display see pic 1 7 O2 Electrode Connector 8 Oximeter body with microcomputer 9 Connector input see pic 4 ...

Page 10: ...Pic 3 Set Oxi hopped wort measurement 1 ICAS 2 4 Legend 1 Outlet hose reduction 4 Intlet hose reduction ...

Page 11: ... 1 CUBE with the program version appears on the display for a moment Then the following text appears memory view 4 Press the key plus There is text memory on the first line and the measured O2 values appear one by one from the oldest to the newest ones on the second line At the same time these values are displayed in SerialWatcher program Windows in column one by one from the oldest value to the n...

Reviews: