T2
If a dust producing appliance is connected
to the extraction system, the sensitivity
must be adjusted for an automatic start of
the extractor.
The following steps are necessary in
order to teach the appliance:
n
The appliance (max. 6A, 120V:600W,
240V:1300W) must be connected to
the socket of the extraction system.
n
Preselect the handpiece speed at
which the extraction system should be
started.
n
Start the appliance or the handpiece
nd keep it running.
n
While the appliance is running, press
and hold the buttons T1 (minus) and
T2 (plus) at the same time for at
least 3 seconds until a signal sounds
and the LEDs of the display L1 light
up. The extraction system will start
running for approx. 3 seconds.
n
Still keep the appliance running.
n
As soon as another signal sounds
and the LEDs of the display L1 start
flashing, switch off the appliance or
stop the handpiece immediately.
n
Wait until a final signal sounds and
the LEDs of the display L1 show the
preselected suction level again.
2.3 Setting up automatic operation