5

Cleaning

Clean the 

SurePower II Battery Pack

 with a soft damp cloth and mild soap. 

Do not immerse or set the 

SurePower II Battery Pack 

in fluids of any kind. 

Do not

 use alcohol 

or ketones to clean the 

SurePower II Battery Pack

.

Maintenance -- Testing and Inspection

Perform a battery check, which verifies that the battery is fully charged and fault free, as part of 
any regularly scheduled unit test. Press the battery’s 

Display

 button to perform the battery 

check.

As part of your defibrillator/monitor maintenance program, verify that the battery’s LEDs work 
(press the 

Display

 button to test LEDs) -- if the LEDs do not work, remove the battery from 

service. Also check the 

SurePower II Battery Pack 

casing for cracks, damaged contacts and 

sockets, or other damage -- if the battery pack is cracked or damaged, remove it from service.

SurePower II Battery Pack 

Icon

LED Indications

This is the 

Recalibration

 icon. 

This LED lights for approximately 10 seconds (after you 
press and release the 

Display

 button) if the battery needs to 

be calibrated. If the 

Recalibration

 LED lights, the runtime 

indicator will not display run time for that battery. For best 
performance of the battery, you should recalibrate the 
battery as soon as possible.

To manually recalibrate the SurePower II Battery Pack, you 
can insert the battery into the SurePower Charger Station 

(using the SurePower

TM

 

II Battery Adaptor) and perform a 

ManualTest

 (for more information, see the 

ZOLL 

SurePower Charger Station 

Operator’s Guide

). After you 

recalibrate the battery, the 

Recalibration 

LED will only flash 

momentarily when you press the 

Display

 button.

This is the 

Fault

 Icon. 

If the SurePower II Battery Pack detects a battery fault, this 
LED lights for approximately 10 seconds after you press the 

Display

 button.

If the 

Fault

 LED lights, the runtime indicator will not display 

run time for that battery.

See the following section for more information on battery 
faults.

Содержание SurePower II

Страница 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Страница 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Страница 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Страница 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Страница 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Страница 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Страница 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Страница 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Страница 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Страница 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...

Отзывы: